RL-TA Series Operation Manual 51
16
Various Functions
16.1
Auto-Zero (Zero Tracking)
This is a function to automatically adjust even a slight deviation of the zero-point caused by the
environment in which the balance is used. In [
3.
A.O
*] of Function 1, the adjustment level can be set.
[
1
] is the lowest setting. The degree of adjustment will increase as the setting increases. Selecting [
O
]
disables this function.
16.2
Auto Backlight OFF
This function turns off the backlight if the balance is left untouched in measurement mode for about three
minutes.
To use the Auto Backlight OFF function, enable it in Function 1 [
A. A.b. 1
].
When the backlight goes off in Auto Backlight OFF mode, touching the balance or pressing any key turns
it back on again.
Under the following conditions, the Auto Backlight OFF function will not work even if [
A. A.b. 1
] is
enabled.
1.
During the setting of Function 1, Function 2, the time or date, or the interval function.
2.
An object is put on the pan, and the display is not stable.
16.3
Auto Power Off
This function turns off the backlight when the balance is left untouched in measurement mode for about
five minutes. This function is only available when the balance is operated by batteries, and is used to
extend the battery lifetime.
To use this function, enable it in Function 1 [
9. A.P. 1
].
[
9. A.P. 1
] does not work under the following conditions:
1.
You are setting Function 1, Function 2, time or date, or the interval function.
2. An object is put on the pan, and the display is not stable.
3.
The balance is operated on the AC adapter.