RL-TA Series Operation Manual 29
11.2
Entering water temperature or the specific gravity of the media
This is for entering correction data for the liquid media to be used for the measurement. Enter water
temperature if the media is water. Otherwise, enter the specific gravity value of the liquid. Use Function
1 [
11.NEd.
*] for selecting the media as follows:
1.
Call the mode to enter a correction value.
Press and hold the Zero/Tare key to enter the temperature
setting mode and then enter the water temperature (for any
media other than water, enter the specific gravity value of
the media).
If the media is water, [
dt
] is displayed on the right side of
the display. Otherwise, [
◄
d
►
] will be indicated instead.
2.
Begin entering a value.
Press the Zero/Tare key.
The [
0
] indication blinks.
3.
Enter the value.
Change the value using the Zero/Tare key. Pressing the
key changes the value as shown on the right.
Press the Function key to enter a higher-order digit.
4.
Save the value.
Upon completing the entry, press the Target key to return to
the weight display mode.
The acceptable range of numeric entry is specified as follows:
Media
Input Data
Range
Water
Water temperature
0.0 to 99.9
°
C
Any liquid other than water
Specific gravity
0.0001 to 9.9999
If any value outside the above specified range is entered, either the minimum or the maximum value within
the range will be set.
The value set is held even after the power is turned off.