RCI-15
65
ZSME300000050
APPENDIX II Function Table
Specification Calibration
Setting Value
Item
Function
Parameter
Description
Default
0
None
1
g
2
Kg
3
t
CSP-01
Unit
4
lb
2
0
None
1
1 Decimal Point
2
2 Decimal Point
CSP-02
Decimal
Point
3
3 Decimal Point
0
1
2
5
10
20
CSP-03
Division
50
Division
1
CSP-04
Max.
Capacity
999999
↓
000000
Max. capacity
999999
CSP-05 Zero range
0 =full range
(
±
1%~30%)
Zero range = calibration zero point
±
(Max.
capacity
×
setting value %)
0
CSP-06
Time of
zero
tracking
0.0 ~ 5.0
(sec)
Time and range of zero tracking should be use
at the same time. If the time is set to 0.0, the
zero tracking function is disabled.
1.0
CSP-07
Range of
zero
tracking
0 ~ 9
Range of zero tracking = (setting value
×
½
)D ,
D=min. division
Range and time of zero tracking should be use
at the same time. If the range is set to 0, the
zero tracking function is disabled.
2
CSP-08
Investigate
time in stable
0.0 ~ 5.0
(sec)
Investigate time and range should be use at the
same time. If the time is set to 0.0, the
investigate time is disabled.
1.0
CSP-09
Investigate
range in
stable
0 ~ 9
Investigate time and range should be use at the
same time. If the range is set to 0, the
investigate range is disabled.
2
0
Action
CSP-10
Weight
unstable,
function
ZERO and
TARE
1
None
0
0
Action
CSP-11
Gross Weight
is negative,
function
TARE
1
None
0