SDM – Slurry Density Meter
41
www.rhosonics.com
CONFIGURATION
SDC625 Map structure (start)
Open Map Dynamic Var Mapping (Online
Device Setup
Detailed setup
Output condition
HART output
Dynamic Var Mapping)
Double click PV is, SV is, Tv is or QV is
PV is setting in map Dynamic Var Mapping
Choose variable for PV is, SV is, TV is and QV is
4.3.2
Configure 4-20 mA Output
This procedure is to specify the 4-20 mA Output.
The PV LRV (Primary Variable Lower Range Value) correspond to 4 mA
The PV URV (Primary Variable Upper Range Value) correspond to 20 mA
The PV Alrm typ (Primary Variable Alarm Type) is the Error reaction mode
The Loop current mode has to be Enabled (Disabled is Multidrop-mode)
The Loop test sets the output to a set value
The D/A trim trims the low and high value on digital level