SDM – Slurry Density Meter
21
www.rhosonics.com
INSTALLATION
2.7.1
USB port
The USB-A connection is located on the front of the SDM.
This USB-A port is used for:
Software update
Collecting Log data
Diagnostics
USB-A port on the front and cable entry cable glands
NOTE:
The cable glands can also be replaced by a Blind plug (standard: M16x1.5), if the entry is not used.
2.7.2
Connection 24 VDC power and HART
The connection for the power and Hart is located in the inside of the SDM. The cable(s) entering
the SDM housing through (a) cable gland(s). Inside the housing the cable(s) are connected to a
screw connector. Below picture shows how to connect the power and HART.
NOTE: Fuse specification
Blow Characteristic:
Fast Acting
Fuse Current:
630mA
Fuse Size:
5mm x 20mm
Voltage Rating VAC:
250V
Cable entry cable gland /stop
(standard: M16x1.5)
USB-A port (front lit removed)