For rear entry connection on 3-10 Litre 
models, we recommend that you use a 
braided flexible hose with a 90

°

 elbow 

for ease of connection. 

 

F.  OVERFLOW / VENT 
 

CONNECTION IMPORTANCE 

Connect a 15 mm (½”) pipe to the 
overflow/vent connection (½”BSP 
nipple). This pipe is to be made of a 
material such that continuous steam will 
not degrade it (i.e. copper). This pipe 
must have a continuous fall, not 
exceeding 3 metres in length and / or 
contain no more than 4 bends (refer to 
figure 2).  

 

 

Figure 1: Water Connection 

 
 

 
 
 
 
 
 

 

 

 

If the site situation requires a pipe length 
exceeding 3 metres, then discharge the 
overflow/drain into a tundish. 
The drain pipe work must not exert a 
pressure of more than 1 metre of head on 
the Lazer boiling unit (refer to figure 2).  

 

During normal operation of the Lazer, the 
overflow/vent connection may discharge 
small quantities of steam and condensate, 
so it is ESSENTIAL that the drain pipe is 
attached to the overflow vent connection. 
This drain pipe must discharge to waste at 
a point where no scald injury, nuisance nor 
inconvenience is caused to people in the 
immediate vicinity. 

 

WARNINGS:  

Ensure the overflow/vent line remains open 
because the Lazer’s tank is not designed to 
be pressurised. Where freezing may occur, 
protect the pipes and install an air break. It 
is recommended to install an air break in 
the overflow/vent drain line, no more than 
300 mm from the Lazer unit. Also it is 
important to connect the overflow/vent and 
water inlet correctly. Incorrect connection of 
pipe work may damage the unit. 

G. DRAIN  

WARNING:  

Before draining the unit, ensure the 
appliance has been switched off, 
and water is not hot enough to 
scald. 
To drain 3-10L models, turn unit off and 
open tap until no water flows out. Remove 
unit from wall and tip over with tap open 
until unit is empty. For 15-40L units. A drain 
plug is accessible from the bottom of the 
unit (note switch unit off before draining) 

IMPORTANT 

Installation and maintenance of the Lazer Boiling Water 

unit shall be carried out by a qualified service person. 

 

Your closest Rheem Service Centre can be contacted by 

telephoning 

131 031

4. 

 

Display Shows

:  “SUN” “7:00 – 7:00”          

“Set OFF hour”.  

  

F) To Service Menu 

 

Select Appropriate Service Mode 

 

Press the (

UP

) button until the desired 

hour appears on the screen and press 
the (

Accept

) button to confirm 

selection. 

i. Error 

Codes 

ii. Boiling 

Temp 

iii. Chiller 

Temp 

iv. Calib 

Reset 

 

v. Software 

Version 

Display Shows

:  “SUN” “7:00 – 7:00”             

“Set OFF Minute”.  

vi. Temp 

Show 

ON/OFF 

 

i) 

Product Error Codes 

Press the (

UP

) button until the desired 

minute appears on the screen and 
press the (

Accept

) button to confirm 

selection and advance to the next day.  

 

Display Shows

:  “MON” “7:30 – 15:30”        

“Set ON hour”. 

   

This function allows easy identification of 
problems occurring with the unit by service 
technicians. 

(Prog>

Service

>Accept>

Error Codes

 

 

ii) 

For Boiling Unit Display Temp mode 

Note:

 Continue with same procedure for 

Monday through to Saturday. 

(Prog>

Service

>Accept>

Boiling Temp

 

Display Shows

:  “XXX ˚C“           

 

 

This function displays the current hot water 
temperature.    

E)  Filter Life 

 

Select the Appropriate Filter Mode 

Press the (

Cancel

) button to revert back to 

the main menu. 

i. Remaining Life

 

ii. Filter Reset

 

 

iii) 

For Chiller Unit Display Temp Mode 

iii. Filter Life

 

 

(NOT USED ON LAZER PRODUCTS) 

i) 

To Select Remaining Life mode 

(Prog>

Service

>Accept>

Chiller Temp

(Prog>

Filter Life

>Accept>

Remaining 

Life

 

Display Shows

:  “XXX ˚C“              

Press the (

Cancel

) button to revert back to 

the main menu. 

 

Display Shows

: “Remaining Life” 

                            “XXXX litres“              

 

iv) 

For Calibration Reset mode 

Press the (

Cancel

) or (

Accept

) button 

to revert back to the main menu. 

(Prog>

Service

>Accept>

Calib.Reset>

Accept) 

 

Display Shows

:  “Calib. Reset“              

 

ii)

 To Select the Filter Reset mode  

This function recalibrates the boiling water 
unit to boiling point.

 

Press the (

Accept

button to recalibrate the boiling point.

 

Press 

(Accept)

 again at the “Are You Sure” 

prompt. Press the (

Cancel

) button to revert 

back to the Main menu.

 

(Prog>

 

Filter Life

 

>Accept>

filter 

reset

>Accept) 

 

Display Shows

:  “Filter Reset”                    

Press the (

Accept

) button to confirm 

selection at which time you are 
prompted “Are you sure?”. Press 
(

Accept

) once more reset the filter and 

to return to the Filter Life Menu. 

 

v) 

To Display the Software Version 

(Prog>

Service

>Accept>

Software 

Version>

Accept) 

 

iii)

 To Select the Filter Life mode  

 

vi) 

Temp Show On/Off  

(Prog>

 Filter Life

 

>Accept>

Set Filter 

Life

>Accept) 

(Prog>

Service

>Accept>

TempShow 

ON/OFF

 

Display Shows

:  “Set Filter Life”                  

“5000 Litres”.  

This function allows the temperature of the 
water within the boiling water unit to be 
displayed permanently on the screen. 

Pressing the (

UP

) button increases the 

5000 litres value in increments of 500 
litres to a maximum value of 12000 
litres. Press the (

Accept

) button to 

confirm selection and the display reverts 
to the Filter menu. 

Press the (

Accept

) button to confirm 

selection and the display reverts to the 
main menu.

 

 

9. 

 

 

Summary of Contents for 10 Litre

Page 1: ...Part No 316937A OWNERS GUIDE INSTALLATION INSTRUCTIONS CLASSIC SERIES POWDER COATED Models 3 5 7 5 10 15 25 40 Litre STAINLESS STEEL Models 7 5 10 15 25 40 Litre IMPORTANT PLEASE LEAVE WITH THE OWNER...

Page 2: ...ction 4 H Tap Outlet 5 J Electrical Requirements 5 4 OPERATION 5 MOUNTING DIMENSION SPECIFICATION FOR ALL MODELS drawing 6 DIMENSION SPECIFICATION FOR ALL MODELS drawing 7 5 FOR UNITS WITH A TIMER 8 A...

Page 3: ...ct your local Rheem office for advice Be sure that the mounting screws are securely inserted into the keyhole slots The screws MUST be anchored in such a way that they will hold the weight of the unit...

Page 4: ...n pipe must discharge to waste at a point where no scald injury nuisance nor inconvenience is caused to people in the immediate vicinity WARNINGS Ensure the overflow vent line remains open because the...

Page 5: ...mise the boiling wate delivery 220 240 Volts AC 50 Hz Single Phase 10 Amps on 3 5 7 5 10 15 litre models 15 Amps on 25 litre model 20 Amps on 40 litre model A flexible cord complete with a plug is sup...

Page 6: ...tre 10 Litre 15 Litre 25 Litre 40 Litre J mm 46 46 46 46 75 75 75 K mm 189 241 241 241 340 340 340 L mm 31 31 31 31 20 20 20 MOUNTING DIMENSION SPECIFICATION All models Figure 2 Overflow Vent Connecti...

Page 7: ...e 25 Litre 40 Litre A mm 283 334 334 334 490 490 490 B mm 143 176 176 176 180 235 325 C mm 400 430 515 615 615 615 615 D mm 234 267 267 267 271 326 416 E mm 15 15 15 15 8 8 8 F mm 50 50 50 50 32 32 32...

Page 8: ...hows STD Pressing the UP button alternates the STD AUTO icons STD operation means that the unit operates 24 hours a day Auto mode reverts the unit to operate at the pre set times on the timer Press th...

Page 9: ...XXX C Press the Cancel button to revert back to the main menu Display Shows Remaining Life XXXX litres iv For Calibration Reset mode Press the Cancel or Accept button to revert back to the main menu...

Page 10: ...ement is properly wired then check its resistance Replace if broken Electronic Controller failure Test the electronic controller 2 The unit fills water to low level and does not heat Thermistor failur...

Page 11: ...NOTES 11...

Page 12: ...re does not cover any plumbing or electrical parts supplied by the installer and not an integral part of the Boiling Water Unit e g pressure limiting valve stop cock electrical switches pumps or fuses...