Level limit switch
Series RF 3000
Technical Information / Instruction manual
page
Push button calibration - Calibration procedure - Uncovered probe only
1. Ensure material level is well
below the probe.
Ensure that the unit is properly mounted and the
material level is
well below the probe
, since the unit
will calibrate to an uncovered probe.
2. Set Sensitivity
Only if required (see page 25)
Set the Sensitivity in the Quickstart Menu, item "D", see page 24.
3. Press CAL button
for 3 seconds
During calibration the display states "CAL", the red LED is blinking.
Wait until calibration is finished (ca. 10 seconds). Then display states the actual measured
capacitance followed by "u" for "Probe
u
ncovered".
Note: If "Code" is displayed, a Lock Code is required. Set the code number with the arrow
buttons and confirm with the Menu button. Then press again the CAL button for 3sec to start
calibration.
If other statements on the display are present, see Troubleshooting, page 37.
Unit is ready to work.
Capacitance
Quickstart
A Capacitance Uncovered probe
B Sensitivity
C Switchpoint
D Capacitance Covered probe
Explanation of calibration procedure:
RFnivo
®
1
2
3
4
5
6
7
8
9
10
11
12
RF 3000 d
gi241014