46

47

Output Specifications for Electrical Muscle Stimulation (EMS):

Waveform

Biphasic

Shape

Square symmetrical with polarity reversal

Maximum Output Voltage (+/-15%) @500Ω Foot: 26 - 35Vp

Body: 18 - 32.6Vp

Pulse Duration (+/-10%):

Foot: 450 or 970µs

Body: 450µs

Net Charge @ 500Ω

[0.001]mC

Maximum Power Density @ 500Ω

Foot: 0.414 mW/cm2

Body: 1.52-3.82 mW/cm² 

ON Time (+/- 10%):

Foot: 1.90 - 8.30s  

Body: 1.90 -33.0s

OFF Time (+/- 10%):

Foot: 1.00 - 1.50s    

Body: 1.00 - 47.0s

The values of PULSE DURATIONS, PULSE repetition frequencies and amplitudes, including 
any d.c. component, shall not deviate by more than ± 20 % when measured with a load 
resistance within the range specified.

If confirmation is required that Revitive works within its Essential Performance after a 
certain period of time, contact the manufacturer

Output Specifications for Transcutaneous Electrical Nerve Stimulation (TENS):

Waveform

Biphasic

Shape

Rectangular

Maximum Output Voltage:      

@ 500Ω

50Vp

Pulse Duration: 

100µs

Net Charge (per pulse @ 500Ω)  

0

μ

C

Maximum Average Power Density (@ 500Ω)

4.19 mW/cm

2

Technical Specifications

The Revitive app allows you to control Revitive from your phone device

Operating Frequencies

 Bluetooth

Operating Range Distance

 Up to 30 metres

Name of product

Revitive Medic Coach

Model

5575AQ

Frequency

EMS: 20Hz – 53Hz

TENS: 80Hz - 130Hz

Output current

EMS: Max 15mA

TENS: Max 19mA

Weight

1.5kg +/- 0.5kg

Dimensions 

360mm (Ø) x 76.5mm (D)

Power consumption

5W

AC adaptor

CE Approved

Power source

Input (adaptor used)

Output

100-240V        AC (     ), 50/60Hz, 0.18A 

5V (      ) DC          ,1.0A

Battery

Lithium Ion, 2600mAh, 3.7V

Battery Life

300 charge cycles

Applied parts

Parts of Revitive that in normal use come into  

physical contact with the user.

Electrode Body Pads - 5cm x 5cm = 25cm

2

Electrode Thigh Pads - 10cm x 12.5cm = 125cm

2

SoftTouch Pads (Foot-Pads) - 253.69cm

2

Durable period (service life) of device and 

Remote Control

4 Years

Bluetooth General Specifications

Standard Compliance

Bluetooth 5.0

Frequency Band

2.402 to 2.480 GHz

Modulation Method

GFSK

5610_IFU02_17816222_PRINT.indd   46-47

5610_IFU02_17816222_PRINT.indd   46-47

21/01/2021   11:12

21/01/2021   11:12

Summary of Contents for Medic Coach

Page 1: ...duct USE ONLY AS DIRECTED If symptoms persist consult your healthcare professional Medic Coach Drug Free Clinically Proven Certified Medical Device User s Manual CIRCULATION BOOSTER 5610_IFU02_1781622...

Page 2: ...What s in the box 4 Parts and Controls 5 6 Introduction to Revitive 7 8 Important Safeguards 9 13 Getting Started 14 26 Using the soft touch foot pads 27 28 Using the Dual Mode Body Pads 29 32 Using...

Page 3: ...booklet Power adaptor Cable length 180cm Revitive Medic Coach Electrode Body Pads Electrode cord Cable length 150cm 1 2 3 4 Parts and controls 70 3o Carry handle Control panel display Power ON OFF Sof...

Page 4: ...hic of gender ethnicity or educational background Electrical Muscle Stimulation EMS Indications for use When using the foot pads Revitive is intended to Reduce pain and discomfort in the legs ankles f...

Page 5: ...your leg below the knee Consult your doctor as soon as possible if you show any signs of the above symptoms Electrical Muscle Stimulation EMS Indications for use continued When using electrode pads ap...

Page 6: ...luding peripherals such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the Revitive System including cables specified by the manufacturer Warnings...

Page 7: ...applying stimulation Be careful when applying stimulation over areas of skin that lack normal sensation It may cause skin irritation due to the inability to feel stimulation until the intensity is to...

Page 8: ...ll charge will take approximately 4 5 hours Once fully charged the full battery indicator will stay illuminated until the power adaptor is unplugged from the device or the electrical socket A full cha...

Page 9: ...to download and install the app from the app store to your phone first Note You will need to be connected to the internet to download the app and complete the setup process Note The Revitive app is pr...

Page 10: ...e asked how you want to open the link in the email choose the Revitive app and just once You will automatically be switched back to the Revitive app If you do not receive an email within one minute re...

Page 11: ...on the Revitive device until you see the Bluetooth symbol flashing on Revitive s control panel This indicates that Revitive is in Bluetooth pairing mode and is ready to connect to your mobile phone F...

Page 12: ...ompleting it you are contributing anonymously to help improve scientific research and product development Tick any that apply and then tap NEXT You will be taken to the Dashboard screen a On the MyThe...

Page 13: ...ive app You can always get back to the Dashboard by pressing This will always be at the bottom of the screen no matter where you are in the app Options menu for further functions and support including...

Page 14: ...he stimulation intensity Tap the large green or buttons to manually decrease or increase the intensity in single units to a comfortable level The intensity level is indicated by the coloured intensity...

Page 15: ...ll ask if you would like Revitive to remember the intensity level you reached for your next session using the Auto Ramp Up feature e The app will now switch to the Dashboard screen You are now ready t...

Page 16: ...e session has ended e Press the electrode pads on your chosen area as shown on page 30 or the Thigh Pads as shown on page 34 gently pressing the adhesive side against the skin Inside of Ankle foot Cal...

Page 17: ...thening program The standard therapy program delivers 10 seconds of stimulation followed by a 20 second rest period which repeats until the end of therapy Delivering 2 muscle contractions per minute G...

Page 18: ...e is disconnected Reconnecting and tapping will re start the stimulation d Remove and store the Thigh Pads by following the instructions on page 32 This section explains how to apply the electrode Thi...

Page 19: ...stimulation will not be possible until the soles of your feet are hydrated In this case it will be necessary to apply a moisturiser Using the SoftTouch foot pads The foot pads deliver EMS to your feet...

Page 20: ...l button on the control panel to start stimulation The Time Display will start to count down in minutes as Revitive cycles through its program Press the to increase the level Press the button to decre...

Page 21: ...ollowing the instructions on page 32 30 Using the Body Pads Revitive comes with one pair of reusable up to 20 30 applications electrode Body Pads which can be used to deliver Electrical Muscle Stimula...

Page 22: ...ys connecting to the other Revitive Remove any pairings to Revitive devices in Settings Bluetooth and take the Revitive that is not required out of Bluetooth range Problem Possible Cause Solution Pair...

Page 23: ...Adjust the positioning of Revitive or lower the intensity level to reduce Revitive tapping Alternatively use a floor mat under Revitive to cushion the sound My legs are aching after treatment You may...

Page 24: ...s Net Charge per pulse 500 0 C Maximum Average Power Density 500 4 19 mW cm2 Technical Specifications The Revitive app allows you to control Revitive from your phone device Operating Frequencies Blue...

Page 25: ...75 700 hPa 1060 hPa Technical Specifications Symbols Complies with European Medical Devices Directive 93 42 EEC UK Conformity Assessed Product conforms to all applicable U K legislative requirements D...

Page 26: ...o retain your original till receipt or invoice to indicate the date of initial purchase To qualify for the 2 year guarantee Revitive must have been used according to the manufacturer s instructions su...

Page 27: ...pyright 2021 Actegy Ltd All rights reserved Actegy Revitive Revitive Medic Coach Circulation Booster OxyWave IsoRocker and SoftTouch Pads are trademarks or registered trademarks of Actegy Ltd The Acte...

Reviews: