Operation and Maintenance Manual – Steam sterilizers mod. BABY S
Rev. 01 del 01-07-2013 - C:\Documenti\Manuali di istruzione\BABY S GB-R01.doc
Pag. 17 di 28
8. MESSAGES DISPLAYED
8.1
Alarm management
MESSAGE
ERROR TYPE
REMEDY
NO POWER
No mains power Fit the machine plug in a single socket that complies with standards
and is without adapters or extensions. Change the fuse and check the
causes. Make sure the master switch (8) is not turned off with closed
door.
NO WATER IN TANK Not enough clean water in load tank Fill up
HIGH
High temperature
Check generator solenoid valve and temperature probe cable (PT100)
LOW
Low temperature
Check solenoid valve, thermostat and heating element of
generator.
Clean the generator probe.
VACUUM Vacuum error Check the vacuum pump and the relevant solenoid valve. Check the vacuum
values on the display unit and on the pressure gauge.
MAX TEMP.
High boiler pressure
Check the temperature probe (PT100) and cable or
detached cable.
PRESS. ERR.
Pressure error Check for any leaks. Check for any heating element malfunctions.
To reset the machine, turn the switch (8) off and on.
8.2 Operating messages
MESSAGE OPERATION IN PROGRESS
Phials
Pre-sterilised phial culture cycle
Print.
Start print
No cycle
Maintenance of instrumental drying
no. cycles
No. of cycle performed by machine
Italian
Current language (set)
End
Cycle finished
8.3 Printed Messages
MESSAGE
OPERATION
Ster. time
Sterilisation time
1
Type of cycle Cycle type
2
Ster. time
Sterilisation time
3
Min.
Cycle min.
4
Sec.
Cycle sec.
5
C
Degrees centigrade
6
Sterilisation
Sterilisation phase
7
Drying
Drying stage
8
End
End of cycle
9
Total time
Total cycle time
10
Test
Test cycle
11
Release
Steam release
12
Stopped
Cycle stopped by operator
13