Operation and Maintenance Manual – Steam sterilizers mod. BABY S
Rev. 01 del 01-07-2013 - C:\Documenti\Manuali di istruzione\BABY S GB-R01.doc
Pag. 14 di 28
7. H O W
T O
A R R A N G E
T H E
I N S T R U M E N T S
F O R
T H E
S T E R I L I Z I N G
P R O C E S S
Before being sterilized, the instruments must be cleaned to remove blood residues, incrustations, foreign
bodies, etc. These impurities can damage the items being sterilized, the autoclave components and
prevent the sterilizing process from being correctly carried out.
These instruments must be made of NO-ferrous material since they could damage or oxidize the other
instruments or the autoclave itself.
To achieve perfect sterilizing, it is therefore important to proceed in the following way:
1. Rinse and brush the instruments under running water immediately after use to remove all deposits.
2. Place the instruments in the ultrasound machine with distilled water and detergent solution if
possible.
3. Thoroughly rinse the instruments in demineralised water to prevent scaling stains.
4. Dry the instruments to eliminate all water residues which could lead to oxidation.
5. Wash, rinse and dry the supplied trays and make sure that they are perfectly clean.
6. To ensure that packaged instruments are perfectly dried, we recommend use of support art. M9050
supplied as an optional.
7. If loose instruments must be sterilized, it is advisable to cover the trays with a paper or fabric napkin
to prevent direct contact with the tray surface: do not place the instruments straight on to the trays.
8. Scissors and forceps should be slightly opened. Mirrors and containers should be placed upside
down for the sterilizing process.
9. It is advisable to keep the instruments separated from each other to achieve a better sterilizing action.
Overloading will impair the sterilizing process.If packaged instruments must be sterilized on the
trays, place the packets with their transparent side downwards and do not overlap. The packets
should not be too large. Drying will be difficult if their edges overlap.
10. The machine can operate with a maximum 3 Kg load.
Sterilizing handpieces:
Carefully read the handpiece operation manual with particular attention to the part about sterilizing
(121°).
Wrap the handpiece in the special sterilizing cloth.
Place the wrapped handpiece on a NON-perforated tray to prevent spilt lubricating oil from polluting
the pipes of the autoclave.
Proceed with a short cycle 121°C sterilizing program.
Remember to immediately remove the handpiece from the autoclave at the end of the sterilizing
process.
For a valid sterilizing process, it is obligatory to add a chemical indicator for each cycle.