MAINTENANCE
CLEANING
Shower sheet and tanks should be thoroughly cleaned after use.
Use liquid cleaning products such as gels and liquid soaps.
Do not use abrasive cleaning products such as scouring cream.
To clean the tanks Reval recommends using BFR 4.0 disinfectant (contact Reval).
NOTE: It is safe to spray BFR 4.0 disinfectant on the heating element.
DESCALING
For the descaling of the stainless steel tanks and the heating element, proceed as follows:
•
Arm yourselves with protective gloves, mask and safety glasses.
•
Remove the metallic grid from the fresh water tank.
•
Spray the scaling product (Actigel réf 1438056) on the walls of the tanks and on the heating element.
•
Let act the product during 2 minutes then rinse the both tanks.
•
Rub the heating element thanks to a sponge to eliminate all tracks of limestone.
•
Rinse abundantly the both tanks.
•
Reposition the grid inside the fresh water tank.
THERMAL DISINFECTION AGAINST LEGIONELLA:
The "Ruby" is compatible with a procedure of thermal disinfection of pipes.
Contact the Reval after-sales department for information.
8/12
Summary of Contents for RUBIS 3017.00
Page 1: ...A4AN3017 doc 23 08 2016 EN Instruction manual SHOWER IN BED SYSTEM RUBIS 3017 00 3017 00 UK...
Page 10: ...MAINTENANCE CLEANING DISINFECTION PROCEDURE BETWEEN 2 SHOWERS average time 5 min 9 12...
Page 11: ...MAINTENANCE CLEANING DISINFECTION PROCEDURE AT THE END OF USE average time 20 min 10 12...