DIRECTIONS FOR USE
USAGE INSTRUCTIONS
1)
Fill the fresh water tank.
2)
Wheel the Ruby Shower in Bed system to the individual.
3)
Connect the electric power cable to an electric socket (16A min required).
4)
Adjust the temperature setting and start heating the water.
5)
Prepare the individual’s bed using the shower sheet.
6)
Shower the individual.
7)
Drain the water from the shower sheet.
8)
Drain the two tanks into a lavatory.
9)
Clean and disinfect the trolley and the shower sheet.
FILLING THE FRESH WATER TANK
Fill the fresh water tank with lukewarm water to the max. level indicated inside the tank.
Do not fill the tank below the mini level indicated inside the tank.
ELECTRICAL CONNECTION
Connect the electric power cable to a 230V electric socket (16A min required), which must be 1m minimum from the
individual’s bed.
WARNING: If the red warning light illuminates it is indicating that the water inside the tank is too warm.
ADJUSTING THE WATER TEMPERATURE
The larger display located at the top of the display shows the real temperature inside
the fresh water tank.
The smaller display located at the bottom of the display shows the set temperature.
To adjust the water temperature, press on the up button during 1 or 2s then adjust
the set temperature thanks to the up & down buttons.
For security reasons the max. temperature cannot exceed 41
°
(as per the factory settings). This temperature and the
controller programming is password protected. To modify these settings please contact Reval technical service.