DustMon RD 100 using PC control unit / software installation
33
4. The "Setup" dialog box opens:
Fig. 21:
"Setup" dialog box
Function
1
Save data at x seconds:
Up to 10 measurement points can be freely selected and inserted here, i.e. the dust
value is, for example, recorded and specified after 5 seconds, 10 seconds etc. after
reaching the MaxValue. These values are shown on the monitor, in the print-out and
are also saved accordingly.
Note: only enter whole numbers.
2
Measurement time [s]
Total duration of the measurement. Typical: 30 seconds
The measurement time can be adjusted to suit requirements/applications and increased
to a maximum 999 seconds. For this the PC must be connected to the DustMon RD
100 and the DustMon RD 100 must be switched on.
If you have set a different measurement time, please click on the "Set/OK" button for
the change to take effect.
3
Result after 30 secs.
If "Result after 30 secs" has been activated (the box has been checked), the DustIndex
will be calculated using the maximum value me value after 30 secs (standard
setting).
If "Result after 30 secs" has been deactivated (the box has not been checked): the
DustIndex will be calculated using the maximum value me value after x secs
(the measurement duration set up beforehand (see (2); e.g. after 120 seconds))
1
2
3
4
5
6
7
8
9