Operating the Device
28
WARNING
W7.0010
Handling of food, pharmaceutical and cosmetic products
Analysed products
Food, pharmaceutical and cosmetic products, which were analysed with
the device must not be consumed, used or circulated.
•
Dispose these substances in accordance with the applicable
regulations.
NOTICE
N13.0007
Range of application of the device
Long-term operation
This laboratory device is designed for eight-hour single-shift operation
with a duty cycle of 30 %.
•
This device may not be used as a production machine nor is it
intended for continuous operation.
NOTICE
N14.0005
Device damage due to liquids
Ingress of liquids into the interior of the device
Mechanical and electronic components are damaged, and the function of
the device is no longer ensured.
•
No wet sieving may be carried out with this device!
6.2 Principle of Operation
The AS 200 tap works with a horizontal circular motion of the sieve stack. At the same time,
vertical tapping impulses are transferred to the sieve stack by means of a tapping arm. With the
circular motion and tapping impulses independent of the mains voltage, the worldwide
comparability of the sieving process is ensured. According to relevant standards, 280
revolutions and 150 impulses per minute are predefined in the Tap Sieve Shaker and cannot be
modified. The sieving time is digitally adjustable.
Summary of Contents for AS200 tap
Page 2: ...2 Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany ...
Page 5: ...Notes on the Manual 5 ...
Page 10: ...Notes on the Manual 10 Service address ...
Page 15: ...Technical Data 15 3 11 Drive Horizontal circular motions 280 rpm Taps 150 min 1 ...
Page 45: ...45 ...
Page 46: ...46 ...
Page 47: ...47 ...
Page 48: ...Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany ...