Cleaning, Wear and Maintenance
37
are thoroughly rinsed with water and dried. The cleaning with strong bases or acids is generally
not recommended.
10.1.1.3
Drying of Test Sieves
Drying ovens of various sizes can be used for drying test sieves (drying temperature < 80 °C).
Additional information concerning ultrasonic cleaning-baths and drying ovens can be found on
the Retsch GmbH homepage (http://www.retsch.com). Also ask for the free expert guide
Sieve
Analysis
– Taking a close look at quality
.
NOTICE
N16.0028
Damage of the sieve mesh fabric
Drying temperature > 80 °C
At higher temperatures, especially fine metal wire meshes can become
warped, leading to a reduced tension of the mesh fabric inside the sieve
frame and hence, makes the test sieve less efficient during the sieving
process.
•
The drying temperature for test sieves must not exceed 80 °C!
10.2 Wear
CAUTION
C10.0013
Personal injury
Improper repairs
This manual does not contain any repair instructions.
•
For safety reasons, repairs may only be carried out by Retsch GmbH
or an authorised representative or by qualified service technicians.
Check the cork stopper (
E
) of the sieve lid, as well as the tapping ram (
KS
) and its O-ring for
wear on a regular basis and replace them if necessary.
Fig. 17:
Tapping ram and cork stopper
E
KS
Summary of Contents for AS200 tap
Page 2: ...2 Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany ...
Page 5: ...Notes on the Manual 5 ...
Page 10: ...Notes on the Manual 10 Service address ...
Page 15: ...Technical Data 15 3 11 Drive Horizontal circular motions 280 rpm Taps 150 min 1 ...
Page 45: ...45 ...
Page 46: ...46 ...
Page 47: ...47 ...
Page 48: ...Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany ...