Packaging, Transport and Installation
17
4.4 Conditions for the Installation Site
Installation height: max. 2
000 m above sea level
Ambient temperature: 5 °C
– 40 °C
NOTICE
N8.0021
Ambient temperature
Temperatures outside the permitted range
Electronic and mechanical components may be damaged.
The performance data alter to an unknown extent.
•
Do not exceed or fall below the permitted temperature range (5 °C to
40 °C ambient temperature) of the device.
Maximum relative humidity < 80 % (at ambient temperatures
≤ 31 °C)
For ambient temperatures U
T
between 31 °C and 40 °C, the maximum relative humidity value L
F
linearly decreases according to L
F
=
–(U
T
– 55) / 0.3:
Ambient temperature
Max. rel. humidity
≤ 31 °C
80 %
33 °C
73.3 %
35 °C
66.7 %
37 °C
60 %
39 °C
53.3 %
40 °C
50 %
NOTICE
N9.0015
Humidity
High relative humidity
Electronic and mechanical components may be damaged.
The performance data alter to an unknown extent.
•
The relative humidity in the vicinity of the device should be kept as
low as possible.
4.5 Electrical Connection
WARNING
W3.0015
Danger to life through electric shock
Connection to sockets without protective earth
An electric shock can cause burns, cardiac arrhythmia, respiratory arrest,
as well as cardiac arrest.
•
The device may only be operated on sockets with protective earth
(PE).
Summary of Contents for AS200 tap
Page 2: ...2 Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany ...
Page 5: ...Notes on the Manual 5 ...
Page 10: ...Notes on the Manual 10 Service address ...
Page 15: ...Technical Data 15 3 11 Drive Horizontal circular motions 280 rpm Taps 150 min 1 ...
Page 45: ...45 ...
Page 46: ...46 ...
Page 47: ...47 ...
Page 48: ...Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany ...