Technical Data
13
3.5 Dimensions and Weight
−
Height without sieve clamping unit:
268 mm
−
Height with sieve clamping unit:
892 mm
−
Width:
540 mm
−
Width with "comfort" clamping unit:
665 mm
−
Depth:
507 mm
−
Weight without sieve stack, without clamping unit: ~ 70 kg
3.6 Required Floor Space
CAUTION
C6.0047
Risk of injury caused by the device falling down
Incorrect installation of the device
−
Due to its weight, the device can cause injuries if it falls down.
•
Only operate the device on a sufficiently large, strong and stable
workstation.
•
Ensure that all of the device feet are securely supported.
−
Width of the base: 570 mm
−
Depth of the base: 570 mm
−
No safety clearances required
Location requirements:
The device must be placed on a vibration-free, plane, stable and free surface to avoid
transmission of vibrations. A level base ensures the uniform distribution of the sample over the
sieve mesh fabric, as well as the stability of the device.
NOTICE
N3.0024
Location requirements
The movement of the sample material causes severe unbalance during
operation
−
Due to the movement of the sample material, the AS 400 control causes
very strong unbalance during operation.
•
The installation must be carried out on a stable, anti-slip, vibration-
free base, which is suitable for both, the weight of the device and the
resulting unbalance during operation.
•
At speeds > 200 U/min, when using test sieves with a diameter of
400 mm and/or with high sieve stacks, the device must be firmly
bolted to the base by means of the transport plates!
•
In order to always ensure safe operation, it is recommended to firmly
screw the AS 400 control to the base regardless of the speed, the
test sieves used and the payload.
3.7 Receptacle Volume
The maximum receptacle volume (the maximum feed quantity) depends on various factors such
as number and aperture size of the test sieves, maximum grain size and width of distribution of
the sample material.