Restek Corporation 

 

 

Reset 
 

To press the reset button use the end of a small tool to access the recessed button. 
 

Single Reset

: Pressing the reset button one time causes the position sensor to be set to zero 

and resets the processor. 
 

Factory Reset

: This requires pressing the reset button while both the + and – buttons are also 

down. The green LED will blink one time and the crimping tool is returned to the factory setting. 
This is useful to find a consistent starting point for adjusting the crimping tool if it is far out of 
adjustment. 

 

 

 
 

Reset 

Summary of Contents for 22358

Page 1: ...Electronic Crimping Tools Models 22358 22359 22360 22362 Operation Guide 1...

Page 2: ...of important data Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met WARNING A WARNING notice warns of a hazard It calls attention to an operating proce...

Page 3: ...p 5 Operation 5 Storage and Shipping 9 Fault Conditions 9 Maintenance Repair 10 This operation manual applies to the following products Model Number 11mm Electronic Crimper 22358 20mm Electronic Crimp...

Page 4: ...ttery Warnings Risk of burns battery may explode or catch fire if mishandled Do not disassemble or dispose of in fire Use only the manufacturer supplied 7 5 volt DC power supply and charge the battery...

Page 5: ...ery The battery must be charged before the crimper or decapper can be used The 7 5 volt DC power supply comes with a set of clips to fit power outlets for most countries Slide the correct clip into th...

Page 6: ...y to engage the motor This switch must be held down until the crimp is complete If the switch is released early the crimper will retract If the Status LED blinks amber after the cycle is complete a fa...

Page 7: ...apper does the rest of the work To adjust the 20mm decapper just make sure that the stroke is long enough to remove the cap When to Recharge the Battery The 6 4 volt battery pack for the electronic cr...

Page 8: ...he position sensor to be set to zero and resets the processor Factory Reset This requires pressing the reset button while both the and buttons are also down The green LED will blink one time and the c...

Page 9: ...ool cycles Early trigger release the tool retracted before completing cycle Try again making sure to hold the button down until the tool is returning to the home position Status 3 Amber Blinks but too...

Page 10: ...Avoid permitting metal parts of the crimping tool come into contact with corrosive material during use If they do try to wipe them clean with a suitable mild neutralizing solution Battery Replacement...

Page 11: ...ormation 11mm decapper leaves caps hanging on vials Decapper adjustment is too low Adjust the decapper to a higher setting by pressing the plus button Jaws are worn or broken The decapper will have to...