Starting therapy
Warmup mode
Warmup mode pre-heats the water before starting therapy so that the air is not cold or dry at the
start of therapy. It is recommended that Warmup mode be run before starting therapy.
Depending on the ambient conditions, the Lumis HFT device will take between 10 to 20 minutes to
warm up. Therapy can be started any time during Warmup mode. After Warmup mode has been
running for 40 minutes, the device will automatically switch to Drying mode.
To run warmup mode:
1.
Select
Run Warmup.
2.
Warmup will take between
10-20 mins. Therapy can be
started at any time during
Warmup
.
3.
When
Warmup
is complete,
Ready
will be displayed.
Start therapy
4.
Fit your high-flow nasal
cannula. See your nasal
cannula user guide for fitting
instructions.
5.
Press the Start/Stop button
to start therapy.
6.
If you have supplemental
oxygen, ensure source is
now turned ON.