Viewing alarms

The AutoSet CS-A is fitted with an alarm module that continuously monitors both therapy and device 

conditions. Alarms are only activated when therapy is running. An alarm condition is indicated by an 

audible sound, a flashing yellow alarm LED and a message on the screen. When the device is turned on, 

the yellow alarm LED will flash, and the alarm will sound to confirm that the alarm is working. 

  

 

9

Alarm activated

 

9

Alarm cleared

 

9

Multiple alarms

When an alarm activates, a 

corresponding alarm message is 

displayed. If multiple alarms are 

active, the latest alarm message 

is displayed and as these are 

cleared each message can be 

read.

To clear an alarm message, 

press 

. This allows you 

to return to the previously 

displayed screen. If the alarm 

condition remains, the alarm will 

re-occur.  

Note: 

Power Fail and Alarm Fail 

alarms are cleared by  

pressing 

.

 

 

To view the list of alarm 

messages, go to the Treatment 

screen and press 

 

until 

Treatment screen 2 is displayed.  

Muting the alarm

To mute an alarm for two minutes, press 

 

once. If the condition remains, the alarm will sound again 

after two minutes. To turn on the alarm again, press 

 

a second time. The alarm LED will remain lit for 

as long as the condition remains.

9

English

Summary of Contents for AutoSet CS-A

Page 1: ...Welcome Guide English AutoSet CS A ADAptive servo ventilatOR...

Page 2: ...include H5i heated humidifier Standard air tubing SlimLine air tubing 3 m air tubing ClimateLine heated air tubing ClimateLineMAX heated air tubing 30W power supply unit does not support H5i Power Sta...

Page 3: ...t one end of the air tubing firmly onto the air outlet 5 Connect the assembled mask system to the free end of the air tubing Notes Always ensure that the AutoSet CS A is placed in an area where the al...

Page 4: ...s you to scroll through the menu and change settings Pushing the dial allows you to enter into a menu and confirm your choice Alarm mute Press once to mute alarms Press a second time to unmute Alarm a...

Page 5: ...oice Using Mask Fit You can use Mask Fit to help you fit your mask properly This feature delivers CPAP pressure for a three minute period prior to starting treatment during which you can check and adj...

Page 6: ...humidification and Climate Control Selecting the mask type To select your type of mask 4 Turn until the mask setting that you require is displayed 5 Press to confirm your choice 6 Push to return to th...

Page 7: ...5 Lie down and arrange the air tubing so that it is free to move if you turn in your sleep 6 To stop treatment at any time remove your mask and or press Notes If your clinician has enabled SmartStart...

Page 8: ...ccessories 9 9 Oximetry data via the oximeter adapter The pressure bar is marked with fixed vertical lines indicating the expiratory and inspiratory pressures While the treatment is ramping indicated...

Page 9: ...can always view your usage hours during the last session and if enabled by your clinician data on Mask Fit and AHI is displayed 9 9 In the Sleep Report only the period can be changed other values are...

Page 10: ...ssage is displayed If multiple alarms are active the latest alarm message is displayed and as these are cleared each message can be read To clear an alarm message press This allows you to return to th...

Page 11: ...sing mild detergent 3 Rinse thoroughly hang and allow to dry 4 Before next use reconnect the air tubing to the air outlet and mask Monthly 1 Wipe the exterior of the device with a damp cloth and mild...

Page 12: ...card 1 Push in the SD card to release it 2 Remove the card 3 Insert the card into the protective folder 4 Send the protective folder back to your clinician as instructed For more information on removi...

Page 13: ...ia See www resmed com for other ResMed locations worldwide For patent information see www resmed com ip S9 AutoSet CS H5i ClimateLine SlimLine and SmartStart are trademarks of ResMed Ltd S9 ClimateLin...

Reviews: