XCell™ C410:V3 Controller
User Guide
66 repligen.com UG-C410v3-06
Table 25. Alarm list
Tag
Description
PLC
Address
Alarm
Enable
Condition Delay
Action
"DB1".ALARM_REG(0)
XCell ATF® Device
Warning condition
has not been reset
within the alarm
delay time, XCell
ATF® Device
function halted.
%DB1.DBX
110.0
Pump is
Running and
Alarm Delay
Timer Setpoint
is Greater than
Zero.
*See Note
1 min
Alarm, System
Stop, Activate
System Fault
Warning Horn
"DB1".ALARM_REG(1)
Main Power Loss
(120/220) while
Pump was running
%DB1.DBX
110.1
First Scan of
PLC on Power
Up.
On first scan
Pump State is
not 0
N/A
Alarm, System
Stop, Activate
System Fault
Warning Horn
"DB1".ALARM_REG(2) P2 Pressure Above
HI Limit
%DB1.DBX
110.2
Pump is
Running
P2 pressure is
greater than
or equal to the
P2 Hi Pressure
alarm
setpoint.
P2 Hi Set Point
Alarm Delay in
10ms
increments.
Alarm, System
Stop, Activate
System Fault
Warning Horn
"DB1".WARN_REG(0)
Flow Set Point
cannot be reached.
Air Supply Regulator
(PRV1) above
maximum pressure
setting
%DB1.DBX
132.0
Pump is
Running
PRV1>=
PRV1MAX
N/A
Warning,
Activate
System Fault
Warning Horn
"DB1".WARN_REG(1)
Flow Set Point
cannot be reached.
Air Supply Regulator
(PRV1) below
minimum pressure
setting
%DB1
.DBX132.1
Pump is
Running
PRV1<=
PRV1MIN
N/A
Warning,
Activate
System Fault
Warning Horn
"DB1".WARN_REG(2)
Exhaust Set Point
cannot be reached.
Vacuum Supply
Regulator (PRV2)
above maximum
vacuum setting
%DB1.DBX
132.2
Pump is
Running
PRV2>=
PRV2MAX
N/A
Warning,
Activate
System Fault
Warning Horn
"DB1".WARN_REG(3)
Exhaust Set Point
cannot be reached.
Vacuum Supply
Regulator (PRV2)
below minimum
vacuum setting
%DB1
.DBX132.3
Pump is
Running
PRV2<=
PRV2MIN
N/A
Warning,
Activate
System Fault
Warning Horn
"DB1".WARN_REG(4)
P2 Pressure below
Lo Limit
%DB1.DBX
132.0
Pump is
Running
P2_PV <=
Pressure
Alarm Lo
Setpoint
P2 Lo Set Point
Alarm Delay in
10ms
increments.
Warning,
Activate
System Fault
Warning Horn
Note 1:
Depends on calibration.