XCell™ C410:V3 Controller
User Guide
16 repligen.com UG-C410v3-06
XCell ATF® Device. In addition to vessel pressure, the hydrostatic pressure generated by the height
difference between vessel liquid level and pump level may assist or hamper the exhaust flow
Pump line (d) -
Located on the P-Box. The line connects the P-Box to the diaphragm pump. A
hydrophobic 0.2 micron filter in this line provides both a sterile barrier and also a potential barrier to
the back flow of liquid from diaphragm pump to controller should a diaphragm rupture. Be sure to
use the hydrophobic filter in the pump line to prevent accidental flow of liquid from the filtration
assembly to the P-Box.
Air pressure regulator
- Typical range of regulator is
0-60 psi
/
4.1bar
. This is a
second stage pressure
regulator for regulating service air inlet pressure to a specified, user required, value.
Typically, the
secondary air pressure is regulated to 35psi/2.4bar. That value is selected because that pressure is
recommended to drive the proportional pressure control valve PRV1 and it is generally the upper
limit of the pressure required to achieve set flow rates.
Air pressure gauge (i) -
Located on the upper side of the P-Box. Typical range of gauge is 0-60psi. It
shows second stage system pressure.
Vacuum gauge (h) -
Located on the upper side of the P-Box. Typical range of gauge is 0 to -30inHg / -
0.95 bar. It shows primary vacuum pressure status.
2.3
Primary electric services
Electric plug (j) -
Located on the E-Box. Electric power (standard 100-240V AC, 50/60Hz).
Power switch (m) -
Located on E-Box.
Power indicator Light -
same as the power switch. Lights green when power is ON.
Stop button (a, n) -
Located on the P-Box and E-Box
.
Either Stop button, causes the system cease operation and default to Standby mode, when the
diaphragm pump defaults to Exhaust
.
2.4
Signal
A total of 4 Sensor input plugs (c)
are provided on the P-Box. All inputs are analogue 4-20 mA. Three
plugs P3, P4 and P5 are for pressure inputs. One plug W1 is for a load cell input.
Ethernet (q) -
communication port for data acquisition on E-Box.
Relays
– Two relay outputs are for relaying alarm conditions (
o
). Two relays for driving a harvest
pump (
p
).
Interconnect (Signal) Cable (I-Cable) -
to relay signal and DC power between E-Box and P-Box.
2.5
C410:V3 Ethernet set-up and connectivity
Ethernet infrastructure
The communication link between the Windows environment and the XCell™ ATF Controller is based
on Profinet® communication protocol. The Profinet® platform is an open Industrial Ethernet
standard developed by the PROFIBUS Organization and is standard Ethernet (IEEE802.3). The
Profinet® communication protocol sends and receives data using the open Ethernet TCP/IP standard
at a bandwidth of 100 Mbit/s and functions identical to traditional industrial Ethernet in providing
“real-time” channel for time-critical communications (i.e. process data).