The LCD/software facility
Step 3
Lock the head in a valid position, A105.0, B-180.0. Qualify the probe
tip on your CMM in the normal manner. The stored tip number should
be noted, together with the corresponding A and B axis positions on
the MIH.
The MIH can now be used in either of the two pre-qualifi ed positions
(A0.0, B0.0 or A105.0, B-180.0) without the need to re-qualify the
probe tip after each move. Simply unlock the head, move to the
required position and lock up. The previously stored tip data for that
head position can be recalled from the CMM computer software ready
to take points.
The head can be used in this way in up to 720 repeatable positions.
The number of positions which are to be used will be dictated by the
workpiece requiring inspection, but in some cases may be restricted
by the number of storage positions for tip data available on your CMM
software. Please consult your CMM supplier for further details.
Attempting to move the head axes whilst locked will cause damage
and may result in movement of the axis position detectors in severe
cases. Should this occur, the MIH software will signal an axis error
message and will ask for a datum reset (fi gure 10 on page 34) when
unlocked.
See step 2 of the datum mode section (6.2) for details of this routine.