Page 1

Calibration

Unsuccessful Calibration

If several consecutive attempts at checking the calibration are unsuccessful:

Loosen or remove the Ocu-Film Tip Cover from the TONO-PEN VET probe tip and 

repeat the calibration check.

If still unsuccessful:

Spray the probe tip with compressed gas, and repeat the calibration check.

If still unsuccessful:

Replace the batteries and repeat the calibration check.

Note: If “CAL” appears followed by a row of four dashes [- - - -], the unit did not go into 

calibration mode.  To correct this, depress the button quicker two times or slower two 

times (see step 2 of calibration procedure). 

Note: Do not take measurements with the instrument if two consecutive “good” calibration 

checks cannot be obtained. The instrument will not take a measurement when the prior 
calibration check was “bAd”.

Note: If the TONO-PEN VET fails to yield a “Good” calibration, call Reichert Technical 

Service.

Summary of Contents for TONO-PEN VET

Page 1: ...User s Guide TONO PEN VET Tonometer Rx Only...

Page 2: ...ontents Introduction 4 Warnings and Cautions 6 Setup 8 Calibration 11 Taking a Measurement 13 Interpreting the LCD 16 Maintenance Storage 17 Troubleshooting 19 Guidance and manufacturer s declaration...

Page 3: ...nstrument the waveform produced by each touch to the anesthetized corneal surface is analyzed and stored for a statistical comparison process Each single valid IOP reading is digitally displayed on th...

Page 4: ...Page Figure 1 TONO PEN VET Components in Storage Case TONO PEN VET Tonometer OCU FILM Tip Covers Stylus Blade Ocu Cel XL Batteries Instruction Manual Introduction...

Page 5: ...O ENSURE THAT CORRECT OPERATION OF THE INSTRUMENT IS MAINTAINED AND TO GUARANTEE THE SAFETY AND RELIABILITY OF THE INSTRUMENT ANY REPAIR OR SERVICE MUST BE PERFORMED BY REICHERT INC OR AN AUTHORIZED S...

Page 6: ...ANTY CAUTION Check to be sure that the batteries are installed correctly Incorrect installation could cause severe damage to the electronics and void the Warranty CAUTION USE OR STORAGE OF THE TONO PE...

Page 7: ...onents are missing or damaged Install the batteries in the battery compartment see page 9 Perform the Calibration Procedure see page 11 Slide a new Ocu Film Tip Cover over the TONO PEN VET unit s stai...

Page 8: ...tion slowly push the stylus forward and pry the cover up to remove Using the other end of the stylus insert the tip into the holes on the underside of the body mold Insert the tip perpendicular into t...

Page 9: ...Page 10 Setup Figure 4 Replacing Batteries...

Page 10: ...etween the first and second button press must be between 0 5 and 1 5 seconds The TONO PEN will beep and display CAL Wait approximately 15 seconds for the TONO PEN to beep The display will change from...

Page 11: ...eplace the batteries and repeat the calibration check Note If CAL appears followed by a row of four dashes the unit did not go into calibration mode To correct this depress the button quicker two time...

Page 12: ...e page 11 Note Calibration should only be performed on the TONO PEN once daily unless otherwise indicated by the instrument The TONO PEN does not require calibration after every patient examination CA...

Page 13: ...f test of the LCD Liquid Crystal Display If any of the LCD segments are not displayed the TONO PEN requires service If a momentary display of CAL is seen followed immediately by a single row of dashes...

Page 14: ...eadings are obtained a final beep will sound and the averaged measurement will appear on the LCD along with the single bar denoting statistical reliability Note If a single row of dashes appears on th...

Page 15: ...ument is activated Two rows of dashes followed by a beep indicates that the instrument is ready to take measurements CAL indicates that the instrument is in the process of calibration check or the ins...

Page 16: ...n the event of unanticipated readings or if a Good calibration check cannot be obtained To clean the probe post Remove the Ocu Film Tip Cover Spray the compressed gas onto the probe tip for approximat...

Page 17: ...ent is used to protect the sensitive electronics against intrusion and should be removed only by the manufacturer Note Removal of the Attachment Screw will void the warranty Specifications Dimensions...

Page 18: ...ion or watch Instructional DVD Old or improperly applied Ocu Film Tip Cover Replace Ocu Film Tip Cover Batteries are low Replace batteries Need to check calibration Perform calibration check procedure...

Page 19: ...tion in a typical commercial or hospital environment Guidance and manufacturer s declaration electromagnetic emissions The TONO PEN VET is intended for use in the electromagnetic environment specified...

Page 20: ...rding to frequency of transmitter meters 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23...

Page 21: ...ed by an electromagnetic site surveys should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80...

Reviews: