WARNING:  NEVER ALLOW LIQUID LEAKING FROM THE BATTERY TO GET IN YOUR EYES OR 

MOUTH.  BECAUSE THIS LIQUID COULD CAUSE SERIOUS PERSONAL INJURY, IF IT DOES 

COME IN CONTACT WITH YOUR EYES, FLUSH THEM IMMEDIATELY WITH PLENTY OF WATER 

AND  CONSULT  A  PHYSICIAN.    LIKEWISE,  IF  THE  LIQUID  GETS  IN  YOUR  MOUTH,  RINSE 

IMMEDIATELY WITH PLENTY OF WATER AND CONSULT A PHYSICIAN.

CAUTION:  AN INSTRUCTION THAT DRAWS ATTENTION TO THE RISK OF DAMAGE TO THE 

PRODUCT.

CAUTION:  DO NOT USE ANY SOLVENTS OR STRONG CLEANING SOLUTIONS ON ANY PART 

OF THIS INSTRUMENT OR DAMAGE TO THE UNIT MAY OCCUR.

CAUTION:  THE TIP OF THE TONO-PEN VET IS A SENSITIVE ELECTRONIC COMPONENT.  BE 

CAREFUL NOT TO BUMP OR JAR THE UNIT AS THIS MAY DAMAGE THE TIP AND EFFECT 

OPERATION.

CAUTION:    DO  NOT  IMMERSE  PROBE  IN  FLUIDS.    THIS  WILL  CAUSE  DAMAGE  TO  THE 

ELECTRONICS AND VOID THE WARRANTY.

CAUTION:  DO NOT STERILIZE THE PROBE.  THIS WILL CAUSE DAMAGE TO THE ELECTRONICS 

AND VOID THE WARRANTY.

CAUTION: CHECK TO BE SURE THAT THE BATTERIES ARE INSTALLED CORRECTLY. 

INCORRECT INSTALLATION COULD CAUSE SEVERE DAMAGE TO THE ELECTRONICS AND 

VOID THE WARRANTY. 

CAUTION: USE OR STORAGE OF THE TONO-PEN VET UNIT WITHOUT AN OCU-FILM TIP COVER 

CAN CAUSE SEVERE DAMAGE TO THE INSTRUMENT AND VOID THE WARRANTY.

CAUTION: REMOVAL OF THE ATTACHMENT SCREW WILL VOID THE WARRANTY.

CAUTION:  MEDICAL ELECTRICAL EQUIPMENT NEEDS SPECIAL PRECAUTIONS REGARDING 

EMC  AND  NEEDS  TO  BE  INSTALLED  AND  PUT  INTO  SERVICE  ACCORDING  TO  THE  EMC 

INFORMATION PROVIDED IN THIS GUIDE.

CAUTION:  ELECTROMAGNETIC INTERFERENCE FROM OTHER DEVICES MAY AFFECT THIS 

INSTRUMENT.  IF INTERFERENCE IS PRESENT, TURN OFF OTHER ELECTRONIC DEVICES, OR 

REMOVE THEM FROM THE IMMEDIATE AREA WHILE OPERATING THIS INSTRUMENT.

CAUTION:  PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT CAN AFFECT MEDICAL 

ELECTRICAL EQUIPMENT.

CAUTION:  THE TONO-PEN VET SHOULD NOT BE USED ADJACENT TO OR STACKED WITH 

OTHER EQUIPMENT.  IF ADJACENT OR STACKED USE IS NECESSARY, THE TONO-PEN VET 

SHOULD BE OBSERVED TO VERIFY NORMAL OPERATION IN THE CONFIGURATION IN WHICH 
IT WILL BE USED.

Warnings and Cautions

Summary of Contents for TONO-PEN VET

Page 1: ...User s Guide TONO PEN VET Tonometer Rx Only...

Page 2: ...ontents Introduction 4 Warnings and Cautions 6 Setup 8 Calibration 11 Taking a Measurement 13 Interpreting the LCD 16 Maintenance Storage 17 Troubleshooting 19 Guidance and manufacturer s declaration...

Page 3: ...nstrument the waveform produced by each touch to the anesthetized corneal surface is analyzed and stored for a statistical comparison process Each single valid IOP reading is digitally displayed on th...

Page 4: ...Page Figure 1 TONO PEN VET Components in Storage Case TONO PEN VET Tonometer OCU FILM Tip Covers Stylus Blade Ocu Cel XL Batteries Instruction Manual Introduction...

Page 5: ...O ENSURE THAT CORRECT OPERATION OF THE INSTRUMENT IS MAINTAINED AND TO GUARANTEE THE SAFETY AND RELIABILITY OF THE INSTRUMENT ANY REPAIR OR SERVICE MUST BE PERFORMED BY REICHERT INC OR AN AUTHORIZED S...

Page 6: ...ANTY CAUTION Check to be sure that the batteries are installed correctly Incorrect installation could cause severe damage to the electronics and void the Warranty CAUTION USE OR STORAGE OF THE TONO PE...

Page 7: ...onents are missing or damaged Install the batteries in the battery compartment see page 9 Perform the Calibration Procedure see page 11 Slide a new Ocu Film Tip Cover over the TONO PEN VET unit s stai...

Page 8: ...tion slowly push the stylus forward and pry the cover up to remove Using the other end of the stylus insert the tip into the holes on the underside of the body mold Insert the tip perpendicular into t...

Page 9: ...Page 10 Setup Figure 4 Replacing Batteries...

Page 10: ...etween the first and second button press must be between 0 5 and 1 5 seconds The TONO PEN will beep and display CAL Wait approximately 15 seconds for the TONO PEN to beep The display will change from...

Page 11: ...eplace the batteries and repeat the calibration check Note If CAL appears followed by a row of four dashes the unit did not go into calibration mode To correct this depress the button quicker two time...

Page 12: ...e page 11 Note Calibration should only be performed on the TONO PEN once daily unless otherwise indicated by the instrument The TONO PEN does not require calibration after every patient examination CA...

Page 13: ...f test of the LCD Liquid Crystal Display If any of the LCD segments are not displayed the TONO PEN requires service If a momentary display of CAL is seen followed immediately by a single row of dashes...

Page 14: ...eadings are obtained a final beep will sound and the averaged measurement will appear on the LCD along with the single bar denoting statistical reliability Note If a single row of dashes appears on th...

Page 15: ...ument is activated Two rows of dashes followed by a beep indicates that the instrument is ready to take measurements CAL indicates that the instrument is in the process of calibration check or the ins...

Page 16: ...n the event of unanticipated readings or if a Good calibration check cannot be obtained To clean the probe post Remove the Ocu Film Tip Cover Spray the compressed gas onto the probe tip for approximat...

Page 17: ...ent is used to protect the sensitive electronics against intrusion and should be removed only by the manufacturer Note Removal of the Attachment Screw will void the warranty Specifications Dimensions...

Page 18: ...ion or watch Instructional DVD Old or improperly applied Ocu Film Tip Cover Replace Ocu Film Tip Cover Batteries are low Replace batteries Need to check calibration Perform calibration check procedure...

Page 19: ...tion in a typical commercial or hospital environment Guidance and manufacturer s declaration electromagnetic emissions The TONO PEN VET is intended for use in the electromagnetic environment specified...

Page 20: ...rding to frequency of transmitter meters 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23...

Page 21: ...ed by an electromagnetic site surveys should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80...

Reviews: