background image

7

Calibration

To properly maintain the accuracy of the Mark III refractometer, calibration of the instrument on a 

routine basis is recommended. Prior to calibration, it is extremely important to completely clean the 

prism surfaces so no sample residue is present. Any residue on the prism surfaces will adversely 

affect the accuracy of the calibration.

 

Calibration options are either Set Point or Span Point. Set Point calibration is performed using 

distilled water as the calibration standard. Reichert recommends performing the Set Point calibration 

daily. Span Point calibration is performed with distilled water and up to 6 additional calibration 

standards.  It  is  not  essential  to  calibrate  at  every  available  point.  Span  Point  calibration  with 

distilled water and at least one calibration standard is recommended to ensure accuracy across 

the reading range of the instrument. Span Point calibration standards can be chosen to closely 

match the reading range of the product being analyzed. 

Set Point Calibration

- Advance to the Calibration, Adjust Set Point menu op  

  tion, and press          .

- Apply distilled water as the calibration standard, as de 

  scribed under Sample Analysis, and allow to equilibrate  

  to the instrument temperature.

- Adjust shadowline to the center of the “X” pattern and press           to 

  calibrate or          to cancel.

- Upon completion the display will revert to the Calibrate menu option screen.

- Press          to return to the reading mode.

Menu Options

Calibrate

   - Adjust Set Point

   - Adjust Span Point

   - Advanced Options

   - Exit 

1.33299

Distilled Water

  

Position shadowline

on center of

“X” in reticle.

Press READ when ready

or SELECT to cancel

Temp

o

 C

Span Point Calibration

The Span Point calibration option allows for up to six additional calibration points. These points 

can be chosen to either calibrate across the entire reading range of the instrument, or to closely 

match a specific range where samples are typically read. The instrument does not have to be 

calibrated at every available point. If you have specific requirements, Reichert offers NIST certi

-

fied calibration fluids as an accessory to this instrument. One bottle of NIST certified calibration 

oil, catalog number 13K41330 is included with the instrument.

Note:

 Prior to performing the Span Point Calibration process, Set Point Calibration must be 

completed.

- Advance to the Calibration, Adjust Span Point menu 

 

  option, and press          .

- Follow the displayed instruction to either activate or 

 

  inactivate a calibration point. Points preceded by * are  

 

  active and those preceded by - are inactive. The choice  

 

  of these points are dictated by which calibration standards  

 

  are available at the time of calibration.

- Proceed with the calibration process by advancing to  

 

  Continue on the display screen and pressing          .

- As the active calibration values are displayed, add that  

  calibration fluid and allow to temperature equilibrate

  Position the shadowline on the center of the “X” pattern  

  and press          to calibrate or          to cancel. Promptly  

  clean the calibration fluid from the instrument to remove  

  all residue.

- The above step will be repeated until all active points are  

  calibrated.

- After the last active point has been calibrated the display  

  will return to the Calibrate option screen.

- Advance to Exit and press          .

- Press          to return to the reading mode.

Menu Options

- Calibrate

   

- Adjust Set Point

   

- Adjust Span Point

   - Advanced Options

   - Exit 

Span Calibration

* 1.40235 

* 1.45746

- 1.48989

- 1.51432

- 1.57230

- 1.63039

Continue

Exit

* = Active

- = Inactive

l l  to choose

SELECT to toggle

Menu Options

- Calibrate

   

- Adjust Set Point

   - Adjust Span Point

   - Advanced Options

   

- Exit

 

1.40235

  

Position shadowline

on center of

“X” in reticle.

Press READ when ready

or SELECT to cancel

Temp

o

 C

Summary of Contents for ABBE MARK III

Page 1: ...ABBE MARK III REFRACTOMETER USER GUIDE...

Page 2: ...instrument with the prism surfaces and instrument clean and dry to prevent premature failure of the prisms and seals due to any harmful effects from sample residue The instrument is programmed to rea...

Page 3: ...Overview 6 Calibration 7 Set Point Calibration 7 Span Point Calibration 7 Editing Calibration Points 8 Operational Parameters Sample Temperature Effects on Readings 8 Automatic Temperature Correction...

Page 4: ...III power input connector or the unit s power cord Damage to the Mark III instrument and or injury to the op erator may occur Warning The Mark III should be used in strict accordance with the instruc...

Page 5: ...47 mm x 47 mm Illumination Six high intensity LED s Prism Assembly Stainless steel assembly with glass prisms Range Channel 1 1 3000 1 7000 nD 0 95 Brix per ICUMSA Readability 0 0001 nD 0 1 Brix Preci...

Page 6: ...Focusing Eyepiece rotate to attain sharp focus of the X pattern and the shadowline Menu Key LCD Display Up Down Keys Select Key Read Key Illumination and Sample Prisms Shadowline Adjustment Knob rota...

Page 7: ...p shadowline In these cases applying the sample with a rubber spatula to squeeze out trapped bubbles or lifting the upper prism sev eral times to remove the trapped air is required The following are t...

Page 8: ...ch as contact lens material that can fit between the prisms when closed Solid samples must be shaped and polished correctly in order to be read The size of solid samples should be no larger than 35mm...

Page 9: ...tion points These points can be chosen to either calibrate across the entire reading range of the instrument or to closely match a specific range where samples are typically read The instrument does n...

Page 10: ...y lower ing the sample temperature will raise the refractive index This change in refractive index due to a change in temperature will cause an appar ent change in concentration unless the reading is...

Page 11: ...me Computer Interface The Mark III includes two serial communications ports Fig 11 Both are DB9 position female connectors that are located on the left side of the instrument Data may be downloaded to...

Page 12: ...products using custom channels These channels represent scales that have been developed to measure a specific product Optional channels can be selected only after they have been uploaded from a PC to...

Page 13: ...ids Percent concentration Brix Solids TC Percent concentration temperature corrected Brix An alternate method of changing modes exists in the reading screen once a sample has been measured All that is...

Page 14: ...s using the keys Time The time may be set in either AM PM or 24 00 00 format Use the or keys to change the displayed value To advance to the next time seqment use the key and the key to exit Date The...

Page 15: ...ider flexibility Any tubing that is highly rigid may apply excessive force on the prism preventing it from closing completely This condition can adversely affect reading accuracy Note If a water bath...

Page 16: ...ff with a suitable solvent to avoid possibly scratching the prism Use caution when removing samples or drying the prism Care and Cleaning Preparing the prism surfaces for samples is essential if consi...

Page 17: ...8152 993530 Accessories Catalog No 13981000 Printer 110 VAC Catalog No 13982000 Printer 220 VAC Catalog No 13107040 901 Null Modem Cable Catalog No 13981000 103 Paper 5 Rolls Catalog No WFAG20047 FUS...

Page 18: ...SA Toll Free 888 849 8955 Phone 716 686 4500 Fax 716 686 4555 email ai reichert com www reichert com Reichert Instruments GmbH Hubertusstrasse 2 D 82229 Seefeld Germany Tel 49 8152 993530 Fax 49 8152...

Reviews: