Error Code
Message Displayed
on main menu
window
Causes
Solutions
R
System Normal
No Error
I
No Signal
Unable to receive signal
Transducers installed
improperly
Loose contact, or not
enough couplant between
transducer and pipe
surface
Pipe liners are too thick,
or the deposition inside
the pipe is too thick.
5. Transducer cables are
not properly connected.
1. Polish the pipe surface and
clean the spot. Remove paint.
2. Make sure there is adequate
couplant.
3. Make sure the transducer is
in tight contact with pipe
surface
4. Check transducer cables
5. Check installation
parameter settings
6. Find a better measurement
site (newer pipe, no corrosion,
no deposition)
J
Hardware Error
Hardware problem
Contact the manufacturer
H
Poor Sig. Detected
Poor signal detected. Similar to
error code I.
Refer to error code I.
E
Current Loop Over
20mA
4-20mA loop output over 120%.
Improper settings for current loop
output
1. Ignore it if current loop
output is not used
2. Check current loop settings
in menu window M32
3. Confirm if the actual flow
rate is too high
Q
Frequency Output
Over
The frequency output is
120% over
2. Improper settings for
frequency output.
3. The actual flow rate is
too high.
1. Ignore if frequency output
is not used
2. Check the values entered in
menu window M35
3. Use a larger value in M35
if needed
4. Confirm if the actual flow
rate is too high
F
System RAM Error
Date Time Error
CPU or IRQ Error
ROM Parity Error
Temporary problems with
RAM, RTC
2. Permanent problems
with hardware
1. Reboot the system
2. Refer to Table 6.1, and
contact the manufacturer
G
Adjusting Gain > s1
Adjusting Gain > s2
Adjusting Gain > s3
Adjusting Gain > s4
(Shown in M311)
**VERIFY***
Instrument is in progress of
adjusting the gain for the signal,
and the number indicates the
progressive steps.
No need for action
K
Empty Pipe
No liquid inside the pipe
1. If the pipe is not full,
relocate the flow meter to
where the pipe is full of
liquid.
Summary of Contents for RH40
Page 62: ...62...
Page 63: ...63...
Page 64: ...64...
Page 65: ...65...
Page 66: ...66...
Page 67: ...67...
Page 68: ...68...
Page 69: ...69...
Page 70: ...70...
Page 71: ...71...
Page 72: ...72...
Page 73: ...73...
Page 74: ...74 8 4 Sound Speed Tables...
Page 75: ...75...
Page 76: ...76...
Page 77: ...77...
Page 78: ...78...
Page 79: ...79...
Page 80: ...80...
Page 81: ...81...
Page 82: ...82...
Page 83: ...83...
Page 84: ...84...
Page 85: ...85...
Page 86: ...86...
Page 87: ...87...
Page 88: ...88...
Page 89: ...89...
Page 90: ...90...