30
§3.3 Transducer Spacing
V-Method Installation
The V-method installation is the most widely used method for daily measurements. The inner diameter of
the pipe ranges from
20mm to 300mm
. It is also known as the
reflective method
. The small sound path
of pipes in this range can interfere with accuracy. This method doubles the sound path, thus, ensuring an
accurate measurement.
Sensors Spacing
TOP VIEW OF PIPE
Z-Method Installation
The Z-method is commonly used for pipe diameters between
300mm and 5,000mm
. Signal strength is
increased with this method.
STOP!
Before mounting, please read Section §3.3!
Transducer spacing is critical to the accuracy of
the meter readings
.
Liquid flow
Ultrasonic signal
Ultrasonic signal
Liquid flow