Recommended separation distances between portable and mobile RF 

communications equipment and the models Curing Light 

The models Curing Light are intended for use in electromagnetic environment in which radiated RF disturbances is 

controlled. The customer or the user of the models Curing Light can help prevent electromagnetic interference by 

maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the 

models Curing Light are recommended below, according to the maximum output power of the communications equipment. 

Rated maximum output 

power  

of transmitter  

Separation distance according to frequency of transmitter 

150kHz to 80MHz 

d=1.2×P1/2 

80MHz to 800MHz 

d=1.2×P1/2 

800MHz to 2.5GHz 

d=2.3×P1/2 

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) 

can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power 

rating of the transmitter in watts (W) accordable to the transmitter manufacturer. 

NOTE  1  At 80 MHz and 800 MHz. the separation distance for the higher frequency range applies.  

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people. 

Summary of Contents for MaxCure5

Page 1: ...Please read this manual before operating RF RCL M014 Edition V2 0 Modify 20210529...

Page 2: ...nting 8 4 Operation 9 5 Cleaning disinfecting and sterilizing 10 6 Transportation storage and maintenance 11 7 Troubleshooting 12 8 Environmental protection 14 9 Statement 14 10 EMC Declaration of con...

Page 3: ...onductors 6 Please recharge the battery in cool and ventilated room 7 It is forbidden to self taking apart the battery in order not to result in short circuit or leakage 8 Keep the Curing Light clean...

Page 4: ...NING No modification of this equipment is allowed WARNING If this equipment is modified appropriate inspection and testing must be conducted to ensure continued safe use of equipment WARNING If the cu...

Page 5: ...against electric shock CLASS II EQUIPMENT According to the degree of protection against electric shock Type B applied part Screw inside outside Used indoor only Recovery Keep dry Handle with care Atmo...

Page 6: ...should be used by trained qualified dentist This Curing Light is used for the principle of ray radiation to solidify the light sensitive resin by shooting at it in a short time it is used to restore t...

Page 7: ...cal shock Class II 2 Protection degree against electrical shock Type B 3 Protection against harmful ingress of water or particular matter ordinary equipment IPX0 can t be waterproof 4 operation mode s...

Page 8: ...art optical fiber Light guide Light Intensity 1000mW cm2 1800mW cm2 1000mW cm2 2500mW cm2 Modes setting TURBO mode Display P1 Illumination 1600mW cm2 1800mW cm2 NORMAL mode Display P2 Illumination 100...

Page 9: ...LED lamp battery and power adapter Table 2 Technical Parameters of built in models Parameters MaxCureG MaxCureH Power supply Input 24VAC 50Hz 60Hz Input 24VAC 50Hz 60Hz Light Intensity 1000mW cm2 2500...

Page 10: ...W 5W Composed mainly main unit LED lamp light hood cable main unit optical fiber light hood LED lamp cable 3 Installation and demounting 3 1 Installation of the table top models a Take off the red cap...

Page 11: ...r is cut off The two power wire should be a little longer than the nylon thread to keep the power wire safe b Take off the red cap from the optical fiber and insert the metal part into the front of th...

Page 12: ...operate the curing light for 10 times under 20s working mode then it will come into over heat protection status And only after 2 minutes sleep it can restart working 200s continuously 5 Cleaning disin...

Page 13: ...Avoid solarization and getting wet in rain or snow during transportation 6 2 Storage 6 2 1 Don t store the machine together with the articles that is combustible poisonous caustic or explosive 6 2 2...

Page 14: ...er 2 The power is off 1 Check the connection of the LED and the power 2 Make sure the power is on Light intensity insufficient 1 Optical fiber isn t inserted well to the bottom 2 The optical fiber has...

Page 15: ...of the optical fiber 1 Reinstall the optical fiber 2 Change a new optical fiber 3 Clear the resin The equipment is not charging when the adapter is connected 1 The adapter is not connected well 2 Faul...

Page 16: ...h EN 60601 1 2 for EMC This does not guarantee in any way that this device will not be effected by electromagnetic interference Avoid using the device in high electromagnetic environment Guidance and...

Page 17: ...such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD lEC 61000 4 2 8 kV contact 2 kV 4 kV 8 kV 15 kV air 8 kV conta...

Page 18: ...power quality should be that of a typical commercial or hospital environment If the user of the models Curing Light require continued operation during power mains interruptions it is recommended that...

Page 19: ...calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 3 5 V1 P1 2 d 1 2 P1 2 80 MHz to 800 MHz d 2 3 P1 2 800 MHz to 2 5 GHz where P is the max...

Page 20: ...F transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the models Curing Light are used exceeds the applicable RF compliance level a...

Page 21: ...smitter W Separation distance according to frequency of transmitter m 150kHz to 80MHz d 1 2 P1 2 80MHz to 800MHz d 1 2 P1 2 800MHz to 2 5GHz d 2 3 P1 2 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2...

Page 22: ...ts of the assembly are cleaned first 2 5 If you find that the reprocessing instructions from the manufacturer seem to be inadequate please inform manufacturer about those inadequacies 2 6 We encourage...

Page 23: ...o rinse thoroughly afterwards at least 3 times with distilled or deionized water 6 CLEANING DISINFECTION 6 1 MECHANICAL REPROCESSING Disinfection must be performed no later than 2 hours after the clea...

Page 24: ...examine the instruments visually 4 Carefully place the products in the disinfection basket Fastening of the products is only permissible if they are freely moveable in the fixture The products are no...

Page 25: ...of other sterilization procedures E g ethylene oxide formaldehyde and low temperature plasma sterilization In such cases please observe the respective valid standards EN IS0 14937 ANSI AAMl IS0 14937...

Page 26: ...td No 8 3 Information Industrial Park High Tech Zone Qixing District Guilin Guangxi 541004 P R China Tel 86 773 7796686 Email refine refine med com Website http www refine med com MedNet EC REP GmbH B...

Page 27: ...ible for any troubles caused by quality problems or products technique and structure 3 The followings are beyond our warranty 1 The damage caused by disobeying the operation instruction or lack of the...

Page 28: ...omer Address Post Code Tel E mail Purchase Date Production Date Distributor Model Product No Guilin Refine Medical Instrument Co Ltd No 8 3 Information Industrial Park High Tech Zone Qixing District G...

Reviews: