15
RF emissions
CISPR11
Class B
The models Curing Light are suitable for used in domestic establishment and in
establishment directly connected to a low voltage power supply network which
supplies buildings used for domestic purposes.
Harmonic emissions
lEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions lEC
61000-3-3
Complies
Guidance & Declaration — electromagnetic immunity
The models Curing Light are intended for use in the electromagnetic environment specified below. The customer or the
user of the models Curing Light should assure that It is used in such an environment.
Immunity test
IEC 60601 test level Compliance level
Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
lEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast transient/
burst
IEC 61000-4-4
±2kV for power supply
lines
±1 kV for Input/output
lines
±2kV for power
supply lines
±1kV for
interconnecting
cable
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
lEC 61000-4-5
±1 kV line to line
±2 kV line to earth
±1 kV line to line
Mains power quality should be that of a typical
commercial or hospital environment.