background image

18

Humidity Calibration Procedure

In order to calibrate the meter humidity calibration standards are required 
(REED R9933, R9975 or R9980). if the meter has been exposed to 
extreme conditions or it has not been used in a long period of time the 
meter should be calibrated by following steps 1 through 5 below.
1.  While in normal %RH measurement mode, hold the 

SET

 and 

LOG

 

buttons simultaneously for approx. 2 seconds to enter the calibration 
mode of %RH.

2.  The LCD display will show the same relative humidity measuring 

value along with the %RH symbol.

3.  Press the   and   buttons to adjust the upper display until it 

matches the %RH Humidity Calibration Standard.

4.  Once the calibration is complete, press "REC" button to save the 

calibration value into the memory which will flash confirming it has 
been saved.

5.  Press the 

NEXT

 button twice to exit the calibration mode and 

resume normal operation. 

Note:

 At any time, you can press the 

ESC

 button to exit the calibration 

mode and resume normal operation.

Data Stream Output

Using Terminal application, a data stream can be viewed from the RS-232 
output, connect RS-232/USB cable between the product and terminal/PC 
and use the following setting to view the data stream.
Terminal Settings:
•  Bits per second: 9600
•  Data bits: 8
•  Parity: none
•  Stop bits: 1

continued...

 

REED Instruments

1-877-849-2127 | [email protected] | www.reedinstruments.com

Summary of Contents for R6050SD

Page 1: ...Data Logging Thermo Hygrometer Instruction Manual R6050SD REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com...

Page 2: ...Thermocouple Measurement Mode 11 Data Hold 11 Recording Maximum and Minimum Readings 11 12 Backlight 12 Setup Mode 12 15 Setting the Time and Date dAtE 13 Setting Data Decimal Format dEC 13 Enabling...

Page 3: ...ort 23 Introduction Thank you for purchasing your REED R6050SD Data Logging Thermo Hygrometer Please read the following instructions carefully before using your instrument By following the steps outli...

Page 4: ...and dew point wet bulb temperatures Contact temperature measurement with a type J K thermocouple option Temperature is measured in C or F Real time data logger with integrated SD memory card User sele...

Page 5: ...1 F 0 5 C 58 F 50 C 0 4 1 8 F 1 C Resolution 0 1 F C Dew Point Measuring Range 13 5 to 120 1 F 25 5 to 48 9 C Accuracy Sum of temp and RH accuracy Resolution 0 1 F C Wet Bulb Measuring Range 6 9 to 12...

Page 6: ...Shut off Yes after 10 minutes off Kick Stand Yes Tripod Mountable Yes Low Battery Indicator Yes Power Supply 6 x AA batteries or AC Adapter optional Data Output Yes RS 232 Battery Life Sampling Time D...

Page 7: ...e K J Thermocouple Input Jack 5 Function Hold Next Button 6 REC Enter Button 7 Log Up Sampling Check Button 8 Humidity Temperature Probe Input Socket 9 RS 232 Output Jack 10 Reset Pin 11 Power Adapter...

Page 8: ...ate Time Values Temperature Measurement Reading 3 Date Time Stamp 4 Temperature Unit of Measure 5 Relative Humidity Unit of Measure 6 Maximum and Minimum Indicator 7 Record Mode Indicator 8 Data Hold...

Page 9: ...erature measurement dp for Dew Point measurement _Et for Wet Bulb measurement and tP for Type K J Thermometer measurement 2 Release the FUNCTION button to confirm selection Humidity Temperature Measur...

Page 10: ...wer part of the display will indicate dP as shown below Wet Bulb Measurement Mode 1 Select the _Et function see Selecting Measurement Modes for details 2 Hold the Probe Handle and face the Probe Head...

Page 11: ...ing a measurement press the HOLD button to freeze the current measurement s on the display 2 While in this mode a HOLD symbol will appear on the display 3 Press the HOLD button again to resume normal...

Page 12: ...tup Mode 2 Press the NEXT button continuously to scroll through the following parameters Parameter Description dAtE Set the time and date dEC Set the decimal format USA 20 00 or European 20 00 PoFF En...

Page 13: ...l point is used to separate units i e 20 00 The European format uses a comma i e 20 00 to separate units To change this setting follow steps 1 and 2 when dEC appears on the LCD 1 Press the and buttons...

Page 14: ...d resume normal operation Selecting the Temperature Unit of Measure t CF Follow steps 1 and 2 when t CF appears on the LCD 1 Press the and buttons to select between C and F 2 Press the ENTER button to...

Page 15: ...e Check button to display the year month day hour and minutes Verify Sampling Rate During normal operation press the Sampling Check button to view the selected sampling rate Data Logging Data Recordin...

Page 16: ...ation Automatic Data Logging Mode In automatic mode the desired data logging sampling rate is set to 1 2 5 10 30 60 120 300 600 1800 or 3600 seconds prior to recording see Setting the Data Logging Sam...

Page 17: ...45 RH 25 Degree C 7 19 07 10 1 13 22 50 RH 35 Degree C 8 19 07 10 1 13 24 60 RH 30 Degree C 9 19 07 10 1 13 26 45 RH 28 Degree C 10 19 07 10 1 13 28 30 RH 22 6 Degree C 11 19 07 10 1 13 30 17 5 RH 22...

Page 18: ...pper display until it matches the RH Humidity Calibration Standard 4 Once the calibration is complete press REC button to save the calibration value into the memory which will flash confirming it has...

Page 19: ...Mono Jack Pin 2 RX Pin 4 DTX Sleeve Tip System Reset If the meter becomes unresponsive or if the display freezes the Reset pin can be used to reset the instrument 1 Use a paper clip or any similar sm...

Page 20: ...ibution HVAC R system installation Greenhouses and industrial spaces Saunas Humidors Museums Lab Monitoring Accessories and Replacement Parts R9933 Humidity Standard 33 R9975 Humidity Standard 75 R998...

Page 21: ...unction by holding down the REC button before attempting to turn the meter off How long can I record for Battery life will depend on a number of factors Quality of Battery Whether the beeper is ON or...

Page 22: ...ew Point measurement _Et for Wet Bulb measurement and tP for Type K J Thermometer measurement Release the FUNCTION button to confirm selection Select the rH option for extended recording of temperatur...

Page 23: ...2127 or info reedinstruments com to discuss the claim and determine the appropriate steps to process the warranty Product Disposal and Recycling Please follow local laws and regulations when disposin...

Page 24: ...Test Measure with Confidence Check out our latest Products REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com...

Reviews: