Setting Overload Alarms 1 & 2
The calibrator allows a user to set overload alarms when the calibrator
temperature is over the temperature set-point. When set and triggered,
will turn off the heating power. When set and triggered,
will turn the cooling fan on.
Note:
The default high overload alarms are set to 5°C (8°F).
These alarms can be set by following the steps below:
1. Press the
button two times until the top screen displays
.
2. Use the
and
arrows to change the temperature value which
will fl ash until confi rmed.
3. Press the
SET
button to confi rm the new high overload alarm 1.
4. Press the
button once to switch to high overload alarm 2 as
indicated by
.
5. Use the
and
arrows to change the temperature value which
will fl ash until confi rmed.
6. Press the
SET
button to confi rm the new high overload alarm 2.
7. Press the
SET
button again to exit set up mode and resume
normal operation.
Enabling/Disabling the Heating Power
1. Press the
button once to display the heating power function as
indicated by
.
2. Use the
and
arrows to select between
(enable
heating power) or
(disable heating power) which will fl ash
until confi rmed.
Note:
The default heating power is set to
.
3. Press the
SET
button to confi rm selection.
4. Press the
SET
button again to exit set up mode and resume
normal operation.
continued...