Turning Off the Calibrator
It is recommended to select a set-point less than 140°F (60°C) and allow
the unit to cool down before turning it off.
Warning:
Do not turn the instrument off at a temperature higher than
212°F (100°C). Lower the temperature below 212°F (100°C)
before powering off.
Selecting Temperature Unit of Measure
When the meter is fi rst powered on the default scale setting is set to Celsius
(°C) and the parameters are locked. In order to change the temperature
scale and unlock parameters, follow the steps below.
1. To unlock, hold down both the
SET
and
buttons simultaneously
and release when the display indicates "At" & "oFF" confi rming that
the parameters are now unlocked.
2. Press the
SET
button for three seconds and release to enter the
setup mode as indicated by
and
.
3. Press button once and the display will indicate
and
which confi rms the current temperature scale.
4. Use the
and
arrows to select between °C and °F.
5. Press the
SET
button to confi rm selection.
6. Press the
SET
button again to exit set up mode and resume
normal operation.
Setting the Temperature Set-Point
1. Use the
and
arrows to increase or decrease the temperature
set-point value.
2. Press the
arrow to increase the temperature value.
3. Press the
arrow to decrease the temperature value.
4. Press the
SET
button to confi rm the new temperature set-point.
5. Press the
SET
button again to exit set up mode and resume
normal operation.