REDLAKE MASD, INC.
TRM P
(see *NOTE)
CR-LF
A positive going edge at the AIA EXPOSE input
trigger connector will start an exposure.
TRM N
(See *NOTE)
CR-LF
A negative going edge at the AIA EXPOSE input or
the trigger connector will start an exposure.
TRM?
Queries the current EXPOSE polarity.
TRE 0
CR-LF
Ends an exposure in controlled mode if TRS SW is
in effect.
TRE 1
CR-LF
Starts an exposure in all modes if TRS SW is in
effect.
TRE?
Queries current state of trigger command.
*
NOTE
: The commands TRM and TRP are interchangeable. TRP may be used in place of TRM.
3.4.13 Transfer Pulse Delay
This command only applies to the double exposure mode. Transfer pulse delay sets the
exposure time for the first image. The exposure time for the second image is 64.946 milliseconds
in 2K x 2K mode, and 32.495 milliseconds with binning set to 1K x 1K.
TYPE IN:
RESPONSE
EXPLANATION
TPD xxx
CR-LF
Sets the delay in increments of 0.2 microseconds.
The argument range is from 1 to 999 microseconds.
If TPD is set to zero, the EXE value controls the
first image exposure time permitting equal exposure
time for both images.
TPD?
Queries the current transfer pulse delay time
setting.
TPD xxx
Means the time is set as indicated.
NOTE
: After TPD xxx or TPD? has been executed, type in “+” to increase the transfer pulse
delay time or type in “-“ to decrease the delay time. The camera will respond with the new
value.
Sample commands with arguments:
TPD 1.6
Sets the delay time to 1.6 microseconds.
TPD 0
The EXE value now sets the delay time.
TPD 999
Sets the delay time to 999 microseconds.
3.4.14 Transfer Pulse Width
The transfer pulse width is fixed at twelve microseconds, except in double exposure mode. This
command sets the duration of the transfer pulse when the camera is in double exposure mode.
The transfer pulse width is the time that the first image has to transfer from the photo diode array
to the CCD array. As the transfer pulse width becomes shorter, the charge transfer between the
arrays does not have time to be completed and image quality will suffer.
TYPE IN:
RESPONSE
EXPLANATION
TPW xx
CR-LF
Sets the delay in increments of one microsecond.
The argument range is from 1 to 10 microseconds.
The default value is 8.
TPW?
Queries the current transfer pulse width setting.
TPW xx
Means the pulse width is set as indicated.
91000116-001 Revision B
3-6
10/15/02
Summary of Contents for MegaPlus ES 4.0/E
Page 6: ...REDLAKE MASD INC THIS PAGE INTENTIONALLY LEFT BLANK 91000116 001 Revision B vi 10 15 02 ...
Page 14: ...REDLAKE MASD INC THIS PAGE INTENTIONALLY LEFT BLANK 91000116 001 Revision B 2 6 10 15 02 ...
Page 38: ...REDLAKE MASD INC THIS PAGE INTENTIONALLY LEFT BLANK 91000116 001 Revision B 4 12 10 15 02 ...
Page 40: ...REDLAKE MASD INC THIS PAGE INTENTIONALLY LEFT BLANK 91000116 001 Revision B II 10 15 02 ...