6.6.5 Installation of calibration adapter
A calibration adapter is available from RBR. This section below will describe how to install this adapter. The full drawing
of the assembly can be found in
Appendix D -Calibration adapter installation
.
Remove anti-biofouling mesh
Follow the instructions for accessing the buffer tube as described in:
Installation of the calibration adapter
Apply a thin layer of grease to the O-ring, see:
Install O-ring into the O-ring groove of the pressure adapter as shown below
Position the calibration adapter over the exposed pressure port of the instrument as shown below
Install the 4 screws and tighten to 1/4 turn past snug. Max 100.00 Nm.