6.3 Installing desiccant
The logger has a dedicated location for placement of desiccant in the battery end cap, which will accommodate three
(3) Dricap desiccant capsules (part #
0003240 (qty9), or
0005668 (qty100)
).
Steps
Remove the battery end cap from the logger - see
Opening and closing the logger
.
Inside the battery end cap, there are 3 recessed sockets. Insert a desiccant capsule into each socket,
ensuring that each is securely placed.
3. Replace battery end cap.
6.4 Twist activation
All loggers are shipped with this feature enabled. This allows the instrument to be enabled without a schedule. To
activate or deactivate the twist feature see
Twist Activation
under the
Gate Condition section of the Ruskin user guide
.
You must first enable this mode and enable the logger. Closing the battery end cap enables the logger. When enabled by
twist activation the logger will vibrate for a few seconds. To stop the logger, open the battery end cap one quarter turn.
When the logger is stopped it will vibrate for three short pulses to provide positive feedback.