Accessories
Pi20
Rev. B 09/2011
35
7.5
Outdoor Enclosure
When the camera is mounted in outdoor environments, the enclosure ensures weatherproofed
installations by providing a high protection rate, a sunshield, and thermostat control heater in cooler
climates.
7.5.1
Technical Data
Scope of delivery
Enclosure with sunshield and Germanium window
Double thermostat control heater
Cable glands
Environment
Rating
IP66/IP67 EN60529 with cable glands
Max. Temperature
-30 to 50°C (-22 to 122°F)
Electrical
Double heater
115/230 VAC, 80 W
Mechanical
Construction
Aluminum
Weight
4.1 kg (9 lb), without camera
Cable entry
M16x1.5
The Germanium window has a transmission factor of 0.87 meaning it transmits 86% of the
radiation before it gets to the camera. For correct temperature readings, that transmission
factor needs to be considered via setting the global emissivity value as a composite
correction factor in the camera’s software. For example, if the target emissivity is 0.93, but
the camera must look through the Germanium window with the transmission factor of
0.87, then the global emissivity setting should be set to 0.81 to compensate for both (0.93 x
0.87 = 0.81).
Figure 16: Outdoor Enclosure
6
6
Photo © 2009, Videotec S.p.A.