Thermalert GP Series Operator’s Manual
51
The length of the air supply hose and the temperature of the cooling air can
be estimated from the chart in Figure B-20.
Note: The “Hose Length” is the length of hose exposed to high ambient temperature
and not the overall length of the hose.
2.1 cubic feet per minute (cfm)
Air Flow: 60 L per minute
1.8 cfm
Air Flow: 40 L per minute
1.4 cfm
Air Flow: 20 L per minute
16
3
100
200
T
amb
[
°
C]
Hose Length
250
150
85
6
9
12
5
1
2
3
4
meters
feet
Figure B-20: Maximum Ambient Temperature Depending on Air
Flow and Hose Length
www.
.com
1.800.561.8187