Profinet
100
MI3
17.2.2.2
Pyrometer Module Parameters
Certain characteristics of same pyrometer modules can be parameterized during the configuration.
Parameter
Description
Setting
Pyrometer number
Set the number of pyrometer
1..8
Emissivity
* 1000 (0.9
900)
100 … 1100
Transmissivity
* 1000 (1.0
1000)
100 … 1000
Averaging time
* 0.1s (1s
10)
0 …9990
Peak hold time
* 0.1s (1s
10)
0 …9990
Valley hold time
* 0.1s (1s
10)
0 …9990
Ambient compensation
Turn ambient compensation
on / off
Ambient temperature
in °C /°F
dev. range min.. max
Setpoint relay
in °C /°F
dev. range min.. max
Relay alarm output control
off,
target temp.,
internal temp.
Laser control
off / on / flashing
Message diagnostics alarm
The diagnostics information of pyrometer is not
transferred to the Profinet I/O controller
message inactive
is transferred to the Profinet I/O controller
message active
Message process alarm
The process alarm of pyrometer is not transferred to
the Profinet I/O controller
message inactive
is transferred to the Profinet I/O controller
message active
Message return of module
The return of pyrometer is not transferred to the
Profinet I/O controller
message inactive
is transferred to the Profinet I/O controller
message active