Appendix
MP150
Protocol
35
Description
Code
Para-
meter
Explanation
Factory
default
Firmware
Revision
bit-1: checksum error in the cali parameter
section
→
%PS
bit-2: checksum error in the temperature
table section
→
%TTS
bit-3: device in warm-up
→
wait some
minutes
bit-4: bias voltage out of range
bit-5: checksum error in the service
parameter section
→
service / may be
ignored
bit-30: no zero-pulse is arriving from the
encoder
–
probably the motor is not
rotating
bit-31: motor is rotating but no data is
arriving at the ad-converters
Get the actual
Current FreQuency
GFQC
-
the FQ
parameter
3.48
Get Identifier
GID
-
Returns scanner identifier
(test SW returns
ID-RAY-M150-000-TEST0-)
-
Get Internal Max.
temperature
GIM
-
Returns the maximal allowed internal
temperature
60
3.22
MAC address
GMAC
-
Read only; returns Hardware (MAC)
address; format: xx:xx:xx:xx:xx:xx, where
xx is hexnumber 00-FF;
returns 00:00:00:00:00:00 if not set
00:1D:8D:xx:xx
:xx
Get Range Bottom GRB
-
Returns min. temp range (°C)
model
dependent
Get Range Full
GRF
-
Returns max. temp range (°C)
model
dependent
Get Range Zoom
GRZ
-
Returns min. temp sub-range (°C)
-
legacy
support:
GRZ = GRF
Get Hold Value
GSH
n
n=1..3: sector/interface
returns current zone value
-
Get Service
Parameter
GSP
t1 t2 T1
T2
t1 = runtime accumulated
t2 = max. runtime without reset
T1 = min. internal temperature
T2 = max. internal temperature
-
Get Sec. Value
GSV
n
n=1..3: sector/interface Returns current
sector value
-
Get Version
Message
GVM
-
Returns software version number
-
Hold Mode
HM
nd
n=1..3: sector/interface d=0: no Hold
Function
d=1: Hold Value for certain time (HT)
d=2: Hold value until trigger
0