Appendix
MP150
Protocol
37
Description
Code
Para-
meter
Explanation
Factory
default
Firmware
Revision
Pixel Mode
PM
d
d = 1...5
pixel per line = 64 * 2
(d
–
1)
(= 64, 128, 256, 512, 1024)
3
(256 pixel)
Pixel Mode
PMX
d x
d = 1...5 (like for PM)
pixel per line = 64 * 2
(d
–
1)
x = 0: omit the surplus pixel
x = 1: calculate the average
x = 2: take the maximum
x = 3: take the minimum
3 0
Reset of the
processor
Reset
-
-
Relay Control
RC
c
c = 0: off
c = 1: on
c = A: auto, controlled by alarm limits
A
always ‘A’ after
reset!
Relay Control
RCX
c s
c = 0: off
c = 1: on
c = A: auto, controlled by alarm limits
s = NO: normally open
s = NC: normally closed
A NO
always ‘A’ after
reset!
3.13
Remote Mode
RM
d
d=’H’: host mode
d=’B’: burst mode
(RMB is defined only for LC001 and ZM00)
Route
(the default
Ethernet gateway)
RO
d.d.d.d
d=0-255
With 0.0.0.0 no route command is sent to
the operating system of the scanner
0.0.0.0
Sector Bottom
SB
ndddd
Temperature minimum of sector
n = 0...3
dddd: temperature in °C
result of GRB
command
Sector Calculation SC
nd
n = 0...3
d = 0: sector calculation off
d = 1: minimum of sector
d = 2: maximum of sector
d = 3: average of sector
d = 4: output of internal temperature
d = 0
nd p
n = 0...3
d = 5: set a fix current
p = current in
μA
nd
d = 6: width measurement
3.41
Background
Temperature
Definition for
Sector Calculation
SCBackgrou
nd
n dcold
dhot
n = 0...9
dcold = cold background temperature to
be excluded from calculation (as integer)
dhot = hot background temperature to be
excluded from calculation (as integer)
0 27000
3.48
Sector Emissivity
SE
nddd
n = 0...3
ddd
= 010...999 → 0.1 ... 0.999
000 → 1.0
0.95
Sector Left
SL
nddd
left starting point of sector n as ddd 1/10° 0
Sector Max
SM
nddd
current maximum in tenth of mA
[15 ... 22 mA]
200