9.1
Troubleshooting
Minor
Problems
Symptom
Probable Cause
Solution
No output
No power to instrument
Check the power supply
Erroneous
temperature
Faulty sensor cable
Verify cable continuity
Erroneous
temperature
Field of view obstruction
Remove the obstruction
Erroneous
temperature
Window lens
Clean the lens
Erroneous
temperature
Wrong emissivity
Correct the setting
Temperature
fluctuates
Wrong signal processing
Correct Peak/Valley Hold or Average
settings
Table
6:
Troubleshooting
60