Operation
6.4.5
Advanced
Valley
Hold
This
function
works
similar
to
the
advanced
peak
hold
function,
except
it
will
search
the
signal
for
a
local
minimum.
6.4.6
Advanced
Peak
Hold
with
Averaging
The
output
signal
delivered
by
the
advanced
peak
hold
functions
tends
to
jump
up
and
down.
This
is
due
to
the
fact,
that
only
maximum
points
of
the
otherwise
homogenous
trace
will
be
shown.
The
user
may
combine
the
functionality
of
the
peak
hold
function
with
the
averaging
function
by
choosing
an
average
time,
thus,
smoothing
the
output
signal
for
convenient
tracing.
output temperature
object temperature
without averaging
Temp
Time
Figure
29:
Advanced
Peak
Hold
with
Averaging
The
advanced
peak
hold
function
with
averaging
is
only
adjustable
by
means
of
the
DataTemp
MultiDrop
Software.
6.4.7
Advanced
Valley
Hold
with
Averaging
This
function
works
similar
to
the
advanced
peak
hold
function
with
averaging,
except
it
will
search
the
signal
for
a
local
minimum.
MID
43