Wuhan Raycus Fiber Laser Technologies Co., Ltd.
User Guide 100W
~
750W Ver: 1.0
19
(2)
A dialog will pop up, to recommend Hyper Terminal as the default telnet program, choose
‘Yes’;
(3)
Hyper terminal will start, type in any name for ‘New Connection’, then click ‘OK’;
(4)
Choose the COM you connected with the RS-232 cable, then click ‘OK’;
(5)
Set up communication protocol: Baud rate-9600, Data Bits-8, Parity-None, Stop bits-1, Flow
Control-None. Then click ‘OK’;
And the hyper terminal setup is finished.
4.7.3
Operation in Hyper Terminal Mode
Before you start the laser, check the wiring, cooling connection and the most important, the
delivery cable and optic terminal are OK. Make sure that the interlock is shorted, and the laser
emission button is up (emission disabled). Then start the chiller.
Turn the key to the ‘ON’ position to start the laser, and the system start self-check, if no error
occurs, the following information will be displayed in the hyper terminal:
Welcome To Use RaycusLaser
RFL-V2.0
Checking RS-232 Connection
RS232 connected.............................
System is doing self-check, please wait.................
Now time :20XX-XX-XX XX:XX:XX
interlock is ok................................. ...
The laser emission is initially disabled...............
Checking temperature...
Module 1 temperature1 = 25 centigrade OK
Module 1 temperature2 = 25 centigrade OK
Module 1 temperature3 = 25 centigrade OK
Module 1 temperature4 = 25 centigrade OK