Wuhan Raycus Fiber Laser Technologies Co., Ltd.
User Guide 100W
~
750W Ver: 1.0
14
6
Remote emission control,
24V level. When the level is
high, the laser emission is enabled, otherwise is
disabled. The function is the same as the ‘LASER’
button on the front panel. ‘LASER’ and the Pin 6 here
can’t work simultaneously. If you use ‘LASER’, left Pin
6 disconnected; if you use Pin 6, keep the ‘LASER’
button up.
24V level, Relative to
PIN 9
7
Control Mode Setting (0->RS232 mode; 1->AD mode)
24V Level
9
GND of Remote Emission Control, Control Mode
Setting and Fault Signal
GND of PIN 6, PIN7
and PIN24
20
Fault Signal Pull up Power
24V Level
22
Analog Voltage Signal
(Laser power control)
0~10V
24
Fault Signal
(High means Laser Fault)
Relative to PIN9
25
GND of Analog Voltage Signal
GND of PIN22
And Figure 17 is the connection diagram
Remote
Emission
Control
Operation
Mode
Setting
Fault
Report
Output
Power
Control
Remote Control Signal
24V or left disconnected
Pull up
0-10V Analog Sinal
Fault Signal Output
GND1
GND1
GND1
GND2
PIN6
PIN9
PIN7
PIN9
PIN20
PIN9
PIN24
PIN22
PIN25
Figure 17: DB25 Connection Diagram