Ratek OM8 Orbital Shaker – User Manual 

-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

 

 

----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- 

Page 10 of 15

 

scrolling)

 

-

 

Push the Start/Stop button to begin.

 

-

 

Press the Start/Stop button once to interrupt the cycle, press again to continue

 

 

To Preset Programs 

-

 

Press Select/Reset to select hours or minutes, then set to the required time

 

-

 

Set the desired speed by pushing the up/down arrows (Single push for single digit adjustments, or hold down for fast 
scrolling)

 

-

 

Hold down the required program button until a “beep” is heard (approx. 4 secs)

 

-

 

The remaining program buttons can be set or changed the same way.

 

-

 

Once set, press the required program button, then press Start/Stop to begin operation.

 

 

 

Adjusting The Counterweight

 

 

 

 

The counterweight is located beneath the platform carrier. It is adjustable and has 4 positions.  The counterweight should be 
set for the expected load to enable the shaker to operate smoothly with maximum balance.  
 
After a short period of use, it should become evident which counterweight positions are most suitable for particular 
applications. 
 

To adjust the counterweight 

-

 

Loosen the 4 retaining screws (front of platform)

 

-

 

Remove the platform by pushing it towards the rear of the cabinet

 

-

 

Locate the counterweight under the platform carrier

 

-

 

To adjust, push the spring loaded latch at the side of the counterweight enabling it to slide freely

 

-

 

Move the weight to the required position then release the latch

 

-

 

Refit the platform, make sure the locating blocks are located correctly with the platform carrier before tightening the 
retaining screws.

 

 

 

Summary of Contents for OM8

Page 1: ...015 English www ratek com au 60 Wadhurst Dve Boronia Victoria 3155 Australia P 613 9887 2161 F 613 9887 2163 E sales ratek com au All Rights Reserved Made in Australia User Manual printed in Australia...

Page 2: ...the product to ensure it has not been damaged in transit Any damage should be reported immediately to the responsible carrier If the product is damaged in any way re pack the product into the supplied...

Page 3: ...TY CONSIDERATIONS 5 5 5 5 Conditions of Operation 5 Safety Labels And Markings 6 OPERATING PROCEDURES 6 6 6 6 Identification of Controls Functions 6 Safety Warnings 7 Safety Recommendations 7 Preparat...

Page 4: ...r the Operator Any item that is not suitably secured to the mixing platform Suitable Environments For The Orbital Shaker The Orbital Shaker is intended for use in a clean laboratory environment only w...

Page 5: ...ng in how to perform their work safely in accordance with any local operational health and safety regulations The operator and all users should be well versed in local emergency procedures as per the...

Page 6: ...as a rotating platform which should not be touched whilst it is moving Keep all body parts away from the moving platform whilst in operation Always stop the platform using the front panel speed contro...

Page 7: ...kplace administrator or electrical maintenance staff to determine if a protective earth power socket is available A surge protected power outlet is strongly recommended as it provides some protection...

Page 8: ...inimum unobstructed distance of 500 millimetres in front of the Orbital Shaker to allow adequate room for the user to maintain a safe operating distance of 300 millimetres CAUTION The Orbital Shaker i...

Page 9: ...erly rated protective earthed wall mounted power supply outlet If there are double adapters or oversized DC power packs causing obstruction of the mains power lead plug these should first be removed E...

Page 10: ...weight The counterweight is located beneath the platform carrier It is adjustable and has 4 positions The counterweight should be set for the expected load to enable the shaker to operate smoothly wit...

Page 11: ...overed bars The bars can be moved horizontally enabling flasks test tube racks and most other types of containers to be secured to the platform TCT25 Tulip Clip Tray This tray is suitable for holding...

Page 12: ...d receptacles are completely dry before restoring electrical power Unplug the equipment from the mains power supply outlet Allow the Orbital Shaker to air dry or use a dry cloth to remove water from a...

Page 13: ...etres Temperature 0 Celsius to 45 Celsius Maximum relative humidity 80 for temperatures up to 31 Celsius decreasing linearly to 50 relative humidity at 40 Celsius MAINS supply voltage fluctuations up...

Page 14: ...a 3 year parts and 12 months labour return to base warranty effective from the date of purchase The product is manufactured in Melbourne Australia The warranty is offered by Ratek Instruments Pty Ltd...

Page 15: ...refund Ratek Instruments will pay your reasonable direct expenses of claiming under this Warranty You may submit details and proof of your expense claim to Ratek Instruments for consideration This Wa...

Reviews: