Fast calibration (orientation)
The
orientation
can
be
started
from
the
Calibration
window
on the
Start
screen
.
●
place
the
calibration
plate
at
the working
distance to the scanner (the first calibration
position),
●
press the
Capture
button,
●
next, following the steps, make two more
shots,
●
after this the accuracy
of
orientation will be
shown. It should be approximately equal to the
accuracy value
of
the
last
calibration.
If the accuracy during orienting starts to be very different from the original one, you need
to re-calibrate the equipment.
Example:
The
initial
accuracy
is
0.024,
accuracy
after
orientation
is
0.065.
The
value
exceeded
0.05,
re-calibration
is
required.
26