Adjustment
R&S
®
RT-ZP03S
8
User Manual 1802.9573.02 ─ 03
3
Adjustment
The probe can be adjusted for low (LF) and high frequency compensation (HF).
Figure 3-1: Compensation trimmer
LF adjustment
1. Connect the probe to a 1 kHz square wave signal.
2. Adjust LF compensation trimmer T1 for optimum square wave response.
Figure 3-2: LF adjustment with 1 kHz signal
RF adjustment
For special measurement tasks, an RF adjustment of the probe can be neces-
sary. Therefore, the probe has two adjustable trimmers below the label beside T1.
Remove the label if you need RF adjustment.
Figure 3-3: Compensation trimmers for RF adjustment