Part 031: Safety requirements for hand-held probe assemblies for electrical mea-
surement and test

Measurement Categories

IEC 61010-2-030 defines measurement categories that rate instruments on their
ability to resist short transient overvoltages that occur in addition to the working
voltage. Use the measurement instrument and accessories only in electrical envi-
ronments for which they are rated.

0 - Instruments without rated measurement category
For measurements performed on circuits not directly connected to mains, for
example, electronics, circuits powered by batteries, and specially protected
secondary circuits. This measurement category is also known as CAT I.

CAT II:
For measurements performed on circuits directly connected to the low-voltage
installation by a standard socket outlet, for example, household appliances
and portable tools.

CAT III:
For measurements performed in the building installation, such as junction
boxes, circuit breakers, distribution boards, and equipment with permanent
connection to the fixed installation.

CAT IV:
For measurements performed at the source of the low-voltage installation,
such as electricity meters and primary overcurrent protection devices.

Disposal

This electronic product is classified within the WEEE/ RoHS category list as
monitoring and control equipment (category 9). Category 9 products are
exempted from the restrictions under the scope of the RoHS directive.

Rohde

 

&

 

Schwarz has developed a disposal concept for the ecofriendly disposal

or recycling of waste material. Rohde

 

&

 

Schwarz fully assumes its obligation as a

producer to take back and dispose of electrical and electronic waste. Contact
your local service representative to dispose of the product.

EC Directives

WEEE Directive 2002/96/EC - Waste Electrical and Electronic Equipment

RoHS Directive 2002/95/EC - Restriction of the use of certain Hazardous Sub-
stances in Electrical and Electronic Equipment

Summary of Contents for RT-ZP03S

Page 1: ...R S RT ZP03S Probe User Manual 1802957302 Version 03 B2 2...

Page 2: ...o rohde schwarz com Internet www rohde schwarz com Subject to change data without tolerance limits is not binding R S is a registered trademark of Rohde Schwarz GmbH Co KG Trade names are trademarks o...

Page 3: ...t is caused by condensation must be expected Temporary condensation occurs only when the product is out of service Using the probe Take the following measures for your safety Before connecting the pro...

Page 4: ...appliances and portable tools CAT III For measurements performed in the building installation such as junction boxes circuit breakers distribution boards and equipment with permanent connection to the...

Page 5: ...3 Meaning of safety labels Safety labels on the product warn against potential hazards Potential hazard Read the product documentation to avoid personal injury or product damage Electrical hazard Indi...

Page 6: ...as Input impedance 10 M II 12 pF meas Input voltage max 400 VRMS 600 V transient overvoltage Table 2 2 Miscellaneous Temperature range 0 C to 40 C Relative humidity max 80 without condensation Altitud...

Page 7: ...nual 1802 9573 02 03 Included in delivery Figure 2 2 Parts included in delivery 1 Retractable hook 2 IC insulating cap 3 Probe 4 Ground lead 5 Adjustment tool 6 Protection cap 7 Identification tags 8...

Page 8: ...e to a 1 kHz square wave signal 2 Adjust LF compensation trimmer T1 for optimum square wave response Figure 3 2 LF adjustment with 1 kHz signal RF adjustment For special measurement tasks an RF adjust...

Page 9: ...n 2 Set the timebase of the oscilloscope to 5 ns div 3 Start the RF adjustment with trimmer T3 Turn it until the peak of the adjusta ble pulse response reaches the end level without any overshoot 4 Tu...

Page 10: ...the product using a soft cloth moistened with either distil led water or isopropyl alcohol Keep in mind that the casing is not waterproof Note Do not use cleaning agents Solvents thinners acetone acid...