Configuration for All Measurement Types
R&S
®
NRX
58
User Manual 1178.5566.02 ─ 05
Source ← Trigger Advanced
See
Slope ← Trigger Advanced
Determines which edge of the envelope power, with internal triggering, or increasing
voltage, with external triggering, is used for triggering.
"Positive"
Rising edge
"Negative"
Falling edge
Remote command:
TRIGger<Measurement>[:CHANnel<Channel>]:SLOPe
Level ← Trigger Advanced
See
Delay ← Trigger Advanced
Sets the delay between the trigger event and the beginning of the actual measurement.
Remote command:
TRIGger<undef>:ALL:DELay[:VALue]
TRIGger<Measurement>[:CHANnel<Channel>]:DELay[:VALue]
Dropout ← Trigger Advanced
If the trigger level has been underrun with internal triggering, it is sometimes useful to
wait some time (drop-out time) before retriggering. The measurement of unwanted sig-
nal components can thus be suppressed. See
Remote command:
TRIGger<Measurement>[:CHANnel<Channel>]:DTIMe
Holdoff ← Trigger Advanced
Sets the hold-off time, see
Chapter 7.3.4, "Hold-Off Time"
Remote command:
TRIGger<Measurement>[:CHANnel<Channel>]:HOLDoff
Hysteresis ← Trigger Advanced
Sets the hysteresis. A trigger event occurs, if the trigger level:
●
Falls below the set value on a rising slope.
●
Rises above the set value on a falling slope.
Thus, you can use this setting to eliminate the effects of noise in the signal for the edge
detector of the trigger system.
Remote command:
TRIGger<Measurement>[:CHANnel<Channel>]:HYSTeresis